Allergy Reliever User Manual AR 1 Series
Content Introduction 2 Parts 3 Warnings 4 Using the Allergy Reliever 5 Battery Information 6 Specification 7 Maintenance and Cautions 8 Explanation of Symbols on Unit 9
Introduction Thank you for choosing the Kinetik Allergy Reliever. This product is a Class IIa medical device that uses technology proven in clinical trials for the treatment of Allergic Rhinitis including some symptoms of Hayfever (seasonal Allergic Rhinitis). Millions of people in the UK suffer from Allergic Rhinitis and/or Hayfever. The suffering is often a lot worse in the summer months due to the presence of pollen and sufferers can become desperate to relieve the symptoms of their allergies.
Parts Battery cover (Open for installing the battery) Connector (Socket) Connector (Plug) Main Body Nose clamp ON/OFF button with indicating light Nasal Probes LED Chips 3
Warnings 1) Children under 16 must be supervised by an adult when using this product. 2) Keep the device out of the reach of children. 3) Not suitable for use with a pacemaker or other implanted medical devices. 4) Not suitable for persons with Nasal Polyposis, sensitive skin in the nose or with poor circulation. 5) Never look directly into the light beams emitted, even when using an optical instrument (for example - glasses). 6) Avoid using the device in the vicinity of flammable or anaesthetic gases.
Using the Allergy Reliever 1) Insert the two nasal probes into the nose as far as it is comfortably possible. 2) Press the ON/OFF button once to turn the unit on. The two wavelengths are activated. 3) One treatment cycle lasts for approximately three minutes. The unit will automatically turn off after the treatment cycle is over. If you need to turn the unit off during treatment, press the ON/OFF button.
Battery Information When the battery power is low, the light on the ON/OFF button will flash to indicate that the battery should be replaced. Open the battery compartment cover by pulling the hook at the back of the unit and insert 2 x AAA batteries (included), observing correct polarity (as shown on the inside of the battery compartment).
Specification Model Number: AR 1 Product Name: Kinetik Allergy Reliever Technology of Treatment: Dual wavelength phototherapy Illuminate Position: Nostrils Wavelengths used: 652 nm (nanometers) & 940 nm (nanometers) Emitting Mode: Continuous Mode Operating Time: Approximately 3 minutes Operating Temperature /Humidity: 15°C ~ 40°C (59°F ~ 104°F) / 30% - 85% RH Storage Temperature /Humidity: -20°C ~ 60°C (-41°F ~ 140°F) / 30% - 85% RH Battery Type: 2 x AAA batteries (included) Battery Life
Maintenance and Cautions 1) Remove the battery after each use to prevent any damage that may be caused by battery deterioration or leakage. 2) This product is not waterproof. Do not clean with any liquid cleaner. To clean the unit, use a damp cloth to wipe the surface. Do not allow any liquid to enter the main unit. 3) Only use Kinetik brand probes and cables with this device. Using non-approved replacement parts may damage the unit and void the warranty. Harvard Medical Devices Ltd.
Explanation of Symbols on Unit Read the instructions. This symbol indicates that this product is a Type BF device. Symbol for "Environment Protection" – Waste electrical products should not be disposed of with household waste. Please recycle where facilities exist. Check with your local Authority or retailer for recycling advice. Symbol for “Manufacturer”. This product complies with MDD93/42/EEC requirements. 0197 Symbol for “European Representative”.