Find us at www.keysight.com
Calibration factor (CF) uncertainty
The typical measurement uncertainties listed are not to be taken as the maximum
CF measurement uncertainties.
Note: The CF uncertainty is dependent on the calibration standard’s uncertainty provided by
Calibration Labs. For the exact uncertainty, refer to the individual sensor calibration report.
U8481A frequency band 25 ºC ± 3 ºC 25 ºC ± 10 ºC 0 ºC to 55 ºC
DC to 10 MHz
1
2.63% 3.30% 3.44%
10 MHz
2
to 30 MHz 1.05% 1.25% 2.35%
>30 MHz to 500 MHz 0.85% 0.89% 1.10%
>500 MHz to 1.2 GHz 0.78% 0.87% 0.87%
>1.2 GHz to 6 GHz 0.91% 1.10% 1.51%
>6 GHz to 14 GHz 1.26% 1.47% 2.04%
>14 GHz to 18 GHz 1.59% 1.96% 2.39%
1. Only applicable for Option 200
2. >10 MHz for Option 200, 10 MHz for Option 100
U8485A frequency band 25 ºC ± 3 ºC 25 ºC ± 10 ºC 0 ºC to 55 ºC
DC to 10 MHz
1
2.37% 2.80% 2.88%
10 MHz
2
to 30 MHz 1.50% 1.49% 2.04%
>30 MHz to 500 MHz 1.37% 1.46% 1.98%
>500 MHz to 1.2 GHz 1.26% 1.52% 2.07%
>1.2 GHz to 6 GHz 1.35% 1.68% 2.40%
>6 GHz to 14 GHz 1.66% 2.26% 2.99%
>14 GHz to 18 GHz 1.83% 2.47% 3.35%
>18 GHz to 26.5 GHz 2.67% 3.75% 4.70%
>26.5 GHz to 33 GHz 3.32% 4.79% 6.41%
1. Only applicable for Option 200
2. >10 MHz for Option 200, 10 MHz for Option 100
U8487A frequency band 25 ºC ± 3 ºC 25 ºC ± 10 ºC 0 ºC to 55 ºC
10 MHz to 30 MHz 1.79% 2.19% 4.15%
>30 MHz to 500 MHz 1.78% 1.90% 2.24%
>500 MHz to 1.2 GHz 1.79% 1.98% 2.34%
>1.2 GHz to 6 GHz 1.82% 2.06% 2.48%
>6 GHz to 14 GHz 1.88% 2.27% 2.53%
>14 GHz to 18 GHz 1.90% 2.36% 2.71%
>18 GHz to 26.5 GHz 2.09% 2.75% 3.23%
>26.5 GHz to 33 GHz 2.66% 3.35% 3.92%
>33 GHz to 34 GHz 2.66% 3.37% 4.10%
>34 GHz to 35 GHz 2.66% 3.39% 4.10%
>35 GHz to 40 GHz 2.66% 4.03% 4.69%
>40 GHz to 45 GHz 3.73% 4.58% 5.43%
>45 GHz to 50 GHz 4.68% 5.71% 6.68%