Datasheet
06 | Keysight | U2741A USB Modular Digital Multimeter - Data Sheet
Frequency accuracy
Function Range
Accuracy
(% of reading + % of
range)
Minimum input
frequency
Temperature
coefficient
(% of range)
Frequency 20 Hz to 300 kHz 0.0200 + 0.003 1 Hz 0.005
Frequency Sensitivity for AC Voltage
Function Range
Minimum sensitivity (RMS sine wave)
Frequency (Hz)
20 ~ 100k 100k ~ 300k
AC voltage
100 mV
2
20 mV 20 mV
1 V 100 mV 120 mV
10 V 1 V 1.2 V
100 V 10 V 20 V
250 V 100 V 120 V
Temperature specifications
Temperature specifications
Function Thermistor type Range Accuracy
Temperature
coefficient
Temperature 5 kΩ thermistor
–80.0°C to 150°C
Probe accuracy + 0.2°C 0.002°C
–112°F to 302°F
Typical Reading Speed (in seconds) Characteristics
Test/Range 100 mV(20 Hz) 1V(20 Hz) 10V(20Hz) 100V(45 Hz) 300V(45 Hz)
ACV 0.979 0.979 0.978 0.979 0.979
Test/Range 10 mA 100 mA 1 A 2 A
ACI 0.979 0.979 0.979 0.979
Freq 1.190
1. Frequency measurement can only be done in auto range mode. Specifications are for 30 minutes warm-up time, using one second aperture.
2. Measuring method is using reciprocal counting technique with AC coupled input at AC voltage function. Gate time of 0.1 second or 1 second.
3. Only applicable for square wave measurement.
Frequency specifications
1
NOTE: To ensure better measurement results and to guard against the change of environment or setup, always enable the Null offset.