Datasheet
04 | Keysight | U2741A USB Modular Digital Multimeter - Data Sheet
DC specifications
1
Function Range Input impedance
Test current/
Burden voltage,
Shunt resistance
Accuracy
±(% of reading + %
of range)
Temperature
coefficient
0°C to 18ºC
28°C to 55°C
Voltage
2
100.000 mV 10 MΩ - 0.015 + 0.008 0.002 + 0.0008
1.00000 V 10 MΩ - 0.015 + 0.005 0.001 + 0.0005
10.0000 V 10 MΩ - 0.018 + 0.005 0.002 + 0.0005
100.000 V 10 MΩ - 0.018 + 0.005 0.002 + 0.0005
300.000 V 10 MΩ - 0.018 + 0.005 0.002 + 0.0005
Current
3
10.0000 mA - < 0.2 V, 10 Ω 0.06 + 0.015 0.005 + 0.0025
100.000 mA - < 0.2 V, 1 Ω 0.06 + 0.005 0.008 + 0.002
1.0000 A - < 0.3 V, 0.1 Ω 0.15 + 0.007 0.005 + 0.002
2.0000 A - < 0.8 V, 0.1 Ω 0.15 + 0.007 0.005 + 0.002
Resistance
4
100.000 Ω - 1.0 mA 0.03 + 0.008 0.006 + 0.0008
1.00000 kΩ - 1.0 mA 0.03 + 0.005 0.006 + 0.0005
10.0000 kΩ - 100 μA 0.03 + 0.005 0.006 + 0.0005
100.000 kΩ - 10.0 μA 0.03 + 0.005 0.006 + 0.0005
1.00000 MΩ - 1 μA 0.06 + 0.005 0.01 + 0.0005
10.0000 MΩ - 225 nA 0.25 + 0.005 0.025 +0.0005
100.000 MΩ - 225 nA || 10 MΩ 2.0 + 0.005 0.3 + 0.0005
Diode test
5
1.0000 V - 1.00 mA 0.015 + 0.03 0.005 + 0.0005
Continuity test
6
1.0000 kΩ - 1.00 mA 0.05 + 0.03 0.005 + 0.0005
1. Specifications are based on 30 minutes warm-up time, NPLC 20 resolution, calibration temperature within 18°C to 28°C, and Null function enabled. For
NPLC 0 and 0.025, add 0.01% of range.
2. 120% over range on all ranges except 300 VDC. Input protection up to 300 VDC.
3. Input protected with externally accessible 2 A, 250 V fast blown fuse.
4. Specifications are applicable with Null function enabled. Otherwise, add 0.2 Ω additional error. Input protection up to 300 VDC. Specifications apply for
NPLC ≥ 1.
5. Specifications are for the voltage measured at the input terminals only.
6. Continuity threshold is fixed at less than 10 Ω.
Product Specifications










