Datasheet
Page 18
Find us at www.keysight.com
Additional Specifications for U2042/44/49/63 and L2061/62/63XA X-Series Peak
and Average Power Sensor (Continued)
U2063 and L2061/62/63XA
Video bandwidth setting
(Normal mode)
LOW MED HIGH OFF
Rise/fall time
2
< 300 MHz 5.3 µs 5.4 µs 1.8 µs 1.8 µs
≥ 300 MHz 0.6 µs 0.64 µs 0.1 µs 0.1 µs
Overshoot
1
< 300 MHz 2% 2% 3% 4%
≥ 300 MHz
12% 15% 9% 5%
The average mode of the X-Series peak and average power sensor provide accurate average
power measurements for broadband modulated signals similar to a thermocouple sensor. This
is due to the X-Series power sensor’ four path diode design, which enables all the diodes to
operate in their square-law region.
1. Specification is based on pulse signal with ≥ 80 ns rise time.
2. Specification is based on pulse signal with 5 ns rise time.
U2042/44/49X A
Video bandwidth setting
(Normal mode)
LOW MED HIGH OFF
Rise/fall time < 300 MHz 6.9 µs 6.9 µs 2.0 µs 2.0 µs
≥ 300 MHz 0.6 µs 0.3 µs 0.1 µs 0.1 µs
Overshoot
1
< 300 MHz 2% 2% 3% 4%
≥ 300 MHz 12% 15% 9% 5%
Characteristic peak flatness
The peak flatness is the flatness of a peak-to-average ratio measurement for various tone
separations of an equal two-tone RF input. Figure 5 below refers to the relative error in
peak-to-average ratio measurements as the tone separation is varied. The measurements were
performed at –10 dBm and applicable for carrier frequency ≥ 300 MHz.