Datasheet

Page 13
Find us at www.keysight.com
Calibration uncertainty
Definition: Uncertainty resulting from non-linearity in the X-Series power sensor detection and correction process. This can be
considered as a combination of traditional linearity, calibration factor and temperature specifications and the uncertainty associated
with the internal calibration process.
Normal mode
Frequency
band
VBW OFF/HIGH VBW MED/LOW VBW OFF/HIGH VBW MED/LOW
U2042XA
1
U2044XA
1
U2049XA U2042XA U2044XA U2049XA L2061XA L2062XA U2063
and
L2063XA
L2061XA L2062XA U2063
and
L2063XA
10 MHz to
30 MHz
5.70% 5.70% 4.50% 4.40% 4.40% 4.50% 4.30% 4.30% 4.40% 4.50% 4.50% 4.30%
> 30 MHz to
500 MHz
5.20% 5.20% 4.10% 3.70% 3.70% 3.90% 3.60% 3.60% 4.10% 3.80% 3.80% 4.00%
> 500 MHz to
1 GHz
5.20% 5.20% 3.90% 3.70% 3.70% 3.90% 3.60% 3.60% 4.10% 3.80% 3.80% 4.00%
> 1 GHz to
6 GHz
5.30% 5.30% 4.00% 3.70% 3.70% 4.00% 3.60% 3.60% 4.10% 3.70% 3.70% 4.00%
> 6 GHz to
10 GHz
5.30% 4.10% 3.70% 4.10% 3.60% 4.10% 3.70% 4.10%
> 10 GHz to
18 GHz
5.40% 4.30% 4.00% 4.20% 3.80% 4.30% 3.80% 4.30%
> 18 GHz to
26.5 GHz
5.00% 4.90% 4.60% 4.50%
> 26.5 GHz to
33 GHz
5.70% 5.60% 5.20% 5.20%
1. Specification valid for environment up to 70% relative humidity. Additional 1.6% to be included for environment up to 95% relative humidity.
Average mode
Frequency band U2041/42XA U2043/44 U2049XA L2051/61XA L2052/62XA U2053/63 and
L2053/63XA
10 MHz to 30 MHz 4.40% 4.40% 4.50% 4.30% 4.30% 4.40%
> 30 MHz to 500 MHz 3.70% 3.70% 3.90% 3.50% 3.50% 3.90%
> 500 MHz to 1 GHz 3.70% 3.70% 3.80% 3.50% 3.50% 3.90%
> 1 GHz to 6 GHz 3.70% 3.70% 3.90% 3.50% 3.50% 3.90%
> 6 GHz to 10 GHz 3.70% 4.00% 3.60% 4.00%
> 10 GHz to 18 GHz 4.00% 4.20% 3.70% 4.20%
> 18 GHz to 26.5 GHz 4.90% 4.50%
> 26.5 GHz to 33 GHz 5.60% 5.10%