Normal mode
Range
1
Zero set (internal) Zero set (external) Zero drift
2
Measurement noise
3
Noise per sample
4
U2000/1/2A
–38 to –15 dBm 47 nW 43 nW 25 nW 28 nW 90 nW
–20 to –6.5 dBm 530 nW 480 nW 230 nW 300 nW 1 µW
–7.5 to –2 dBm 30 µW 27 µW 19 µW 20 µW 55 µW
–4 to 15 dBm 32 µW 30 µW 24 µW 21 µW 85 µW
10 to 20 dBm 270 µW 200 µW 110 µW 180 µW 550 µW
U2000/1/2H
–28 to –5 dBm 730 nW 500 nW 300 nW 310 nW 900 nW
–10 to 3.5 dBm 5.3 µW 4.8 µW 3 µW 5 µW 10 µW
–2.5 to 8 dBm 330 µW 270 µW 190 µW 230 µW 550 µW
8 to 25 dBm 440 µW 300 µW 300 µW 260 µW 850 µW
20 to 30 dBm 3.9 µW 2.8 mW 1.1 mW 2.8 mW 5.5 mW
U2000/1B
–8 to 15 dBm 47 nW 43 µW 25 µW 28 µW 90 µW
10 to 23.5 dBm 530 nW 480 µW 230 µW 300 µW 1 mW
22.5 to 28 dBm 30 µW 27 mW 19 mW 20 mW 55 mW
26 to 44 dBm 32 µW 34 mW 24 mW 21 mW 85 mW
1. Condition: (i) 0 to 55 °C and (ii) 95% relative humidity at 40 °C non-condensing.
2. Within one hour after zero set, at a constant temperature, after a 24-hour warm-up of the power sensor.
3. The number of averages at 1 for Normal speed, gate length of 2.27 ms, measured over one minute interval and two standard deviations.
4. The Noise Per Sample specification is only applicable for gated power working range stated in the “Normal Mode Key Specifications and Characteristics”
table.