Datasheet
Electrical Specifications for U1213A (continued)
AC current
Range Resolution
Accuracy *N1
Maximum overload45 ~ 65 Hz 65 ~ 200 Hz 200 ~ 300 Hz 300 ~ 400 Hz
40 A 0.01 A 2.0 % + 10 3.0 % + 10 3.5 % + 10 6.5 % + 10
1000 A R.M.S.400 A 0.1 A 2.0 % + 5 3.0 % + 5 3.5 % + 5 6.5 % + 5
1000 A 1 A 2.0 % + 5 3.0 % + 5 3.5 % + 5 6.5 % + 5
– N1: The maximum verification of current and frequency product is less than 400,000 A x Hz
AC+DC current
Range Resolution
Accuracy *N1
Maximum overload
45 ~ 65 Hz 65 ~ 200 Hz 200 ~ 300 Hz 300 ~ 400 Hz
40 A 0.01 A 3.5 % + 25 4.5 % + 25 5.0 % + 25 8.0 % + 25
1000 A R.M.S.400 A 0.1 A 3.5 % + 9 4.5 % + 9 5.0 % + 9 8.0 % + 9
1000 A 1 A 4.5 % + 9 5.0 % + 9 5.5 % + 9 8.5 % + 9
Current (1 ms peak hold)
Range Resolution Accuracy Maximum overload
40 A 0.01 A 2.0 % + 70
1000 A R.M.S.400 A 0.1 A 2.0 % + 43
1000 A 1 A 2.0 % + 43
– Specified accuracy for changes > 1 ms in duration
Temperature test
Thermal type Range Resolution Accuracy
K
–200 ~ –40 °C 0.1 °C 1 % + 3 °C
–40 ~ 1372 °C 0.1 °C 1 % + 1 °C
–328 ~ –40 °F 0.1 °F 1 % + 6 °F
–40 ~ 2502 °F 0.1 °F 1 % + 2 °F
Notes:
1. The accuracy does not include the tolerance of the thermocouple probe, and the meter must be operational for at least one hour
2. Do not allow the temperature sensor to contact a surface that is energized above 33 V R.M.S. or 70 V DC. Such voltages will pose a shock
hazard.
3. The temperature calculation is according to the standard of EN/IEC-60548-1 and NIST175