Datasheet
9
Frequency Sensitivity Specications
For voltage measurements
Input range Minimum sensitivity (rms sine wave)
Maximum input for specied accuracy
10 Hz to 10 kHz 10 to 60 kHz
U1192/3/4A U1192/3/4A
60 V 6.0 V 30 V
600 V 60 V -
For current measurements
Input range Minimum sensitivity (rms sine wave)
Maximum input for specied accuracy
45 Hz to 1 kHz
U1192/3/4A
60 A 6.0 A
600 A 60 A
Continuity specications
Range Resolution
Accuracy
U1191A U1192A U1193A U1194A Test current
600 Ω 0.1 Ω 0.8% + 5 0.8% + 5 0.8% + 5 0.8% + 3 0.1 mA
Notes for continuity specifications:
1. Overload protection: 600 Vrms for short circuits with < 0.1 mA current.
2. Maximum open voltage is 1.4 V.
3. Built-in buzzer beeps continuously when the reading measured is less than 30 Ω and does not beep when the measured resistance is more
than 200 Ω. Buzzer may either sound or not between 30 Ω and 200 Ω.
4. Continuity indicator: 2.7 kHz tone buzzer.
Measuring rate (approximate)
Function
Times/second
U1191A U1192A U1193A U1194A
AC V 3 3 3 3
DC V 3 3 3 3
Ω 2 2 2 2
Diode 3 3 3 3
Capacitance — 2 times/second for 600 µF
1 time/9 seconds for 6 mF
2 times/second for 600 µF
1 time/9 seconds for 6 mF
2 times/second for 600 µF
1 time/9 seconds for 6 mF
Temperature — — — 2
DC A — — — 3
AC A 3 3 3 3
Frequency — 3 (> 10 Hz) 3 (> 10 Hz) 3 (> 10 Hz)