Datasheet
09 | Keysight | N8700 Series System DC Power Supplies - Data Sheet
Performance Specications
1. Minimum voltage is guaranteed to maximum 0.2% of rated output voltage.
2. Minimum current is guaranteed to maximum 0.4% of rated output current.
3. 20 MHz
4. 5 Hz - 1 MHz
5. From no-load to full-load, constant input voltage. Maximum drop in remote sense.
6. For load voltage change equal to the unit voltage rating, constant input voltage
7. Single-phase and 3-Phase 208 V models: 170~265 VAC, constant load. 3-Phase 400 V models: 342~460 VAC, constant load.
8. The constant current programming readback and monitoring accuracy does not include the warm-up and load regulation thermal drift.
9. Time for output voltage to recover within 0.5% of its rated output for a load change 10 - 90% of rated output current, local sense.
N8754A N8755A N8756A N8757A N8758A
DC output ratings
Voltage
1
20 V 30 V 40 V 60 V 80 V
Current
2
250 A 170 A 125 A 85 A 65 A
Power 5000 W 5100 W 5000 W 5100 W 5200 W
Output ripple and noise
CV
p-p
3
75 mV 75 mV 75 mV 75 mV 100 mV
CV
rms
4
10 mV 10 mV 10 mV 10 mV 15 mV
Load effect
CV load regulation
5
8 mV 9.5 mV 11 mV 14 mV 17 mV
CC load regulation
6
250 mA 170 mA 125 mA 85 mA 65 mA
Source effect
CV line regulation
7
2 mV 3 mV 4 mV 6 mV 8 mV
CC line regulation
7
125 mA 85 mA 62.5 mA 42.5 mA 32.5 mA
Programming accuracy
Voltage
1
0.025% + 15 mV 22.5 mV 30 mV 45 mV 60 mV
Current
2, 8
0.1% +
750 mA 510 mA 375 mA 255 mA 195 mA
Measurement accuracy
Voltage 0.025% + 25 mV 37.5 mV 50 mV 75 mV 100 mV
Current
8
0.1% +
750 mA 510 mA 375 mA 255 mA 195 mA
Load transient recovery time
Time
9
<1 ms <1 ms <1 ms <1 ms <1 ms