Performance Characteristics (continued)
Settling time
1
Manual filter, 10-dB decreasing power step for normal and x2 modes (not across switching point)
Number of averages 1 2 4 8 16 32 64 128 256 512 1024
Settling time (s)
Normal speed 0.15 0.2 0.3 0.5 1.1 1.9 3.4 6.6 13 27 57
x2 speed 0.15 0.18 0.22 0.35 0.55 1.1 1.9 3.5 6.9 14.5 33
1. Settling time: 0 to 99% settled readings over the GPIB
Typical
settling
times
x2 mode Normal mode
150 ms 150 ms
180 ms 200 ms
400 ms 1 s
3.6 s 6.6 s
6.6 s 13.5 s
20 dB
10 dB
10 dB
10 dB
5 dB
Sensor
dynamic
range
Maximum dBm
Minimum dBm
Auto filter, default resolution, 10-dB decreasing power step for normal and x2 modes
(not across switching point)
Calibration factor and reflection coefficient
Calibration factor (CF) and reflection coefficient (Rho) data are unique to each sensor.
The CF corrects for the frequency response of the sensor. The Keysight EPM Series,
EPM-P Series, and P-Series power meters automatically read the CF data stored in the
sensor’s EEPROM and use it to make the corrections.
The reflection coefficient (Rho or ρ) relates to the SWR according to the following formula:
SWR =
1 + ρ
1 – ρ
Typical measurement uncertainties of the calibration factor (CF) are listed in the following
table. There is only one set of CF data used for both the high and low range of each sensor.
The typical measurement uncertainty data listed in the table is meant to help users on
the measurement uncertainty estimation. These values are only a guideline and are not to
be used in any accurate uncertainty calculations. For accurate measurement uncertainty
values, please refer to the measurement report
1
of the specific sensor.
1. Only applicable with the purchase of Option 1A7 or Option A6J