Datasheet
Page 10
Find us at www.keysight.com
Performance Characteristics (continued)
Switching point
The N8480 Series power sensors (excluding Option CFT) have two measurement ranges:
a lower power range and an upper power range. The power meter automatically selects
the proper power level range. To avoid unnecessary switching when the power level is
near the switch point, switching point hysteresis has been added.
N8481A/82A/85A/87A/88A and N8486AQ/AR power sensors
This hysteresis causes the lower power range to remain selected until approximately
–0.5 dBm as the power level is increased. Above this power, the upper power range is
selected.
Likewise, the upper power range remains selected until approximately –1.5 dBm as the
signal level decreases. Below this power, the lower power range is selected.
N8481B/82B power sensors
This hysteresis causes the lower power range to remain selected until approximately
29.5 dBm as the power level is increased. Above this power, the upper power range is
selected.
Likewise, the upper power range remains selected until approximately 28.5 dBm as the
signal level decreases. Below this power, the lower power range is selected.
N8481H/82H power sensors
This hysteresis causes the lower power range to remain selected until approximately
17.5 dBm as the power level is increased. Above this power, the upper power range is
selected.
Likewise, the upper power range remains selected until approximately 16.5 dBm as the
signal level decreases. Below this power, the lower power range is selected.
Power range in range setting
Sensor Range setting Lower range Upper range
N8481A/82A/85A/87A/88A and
N8486AQ/AR excluding Option CFT
AUTO (default) –35 to –1 dBm –1 to +20 dBm
LOWER –35 to –1 dBm —
UPPER
1
— –30 to +20 dBm
N8481B/82B excluding Option CFT AUTO (default) –5 to +29 dBm +29 to +44 dBm
LOWER –5 to +29 dBm —
UPPER
1
— 0 to +44 dBm
N8481H/82H excluding Option CFT AUTO (default) –15 to +17 dBm +17 to +35 dBm
LOWER –15 to +17 dBm —
UPPER
1
— –10 to +35 dBm
1. Recommended for pulse signals measurement with period of more than one second