Datasheet
N6731B-N6792A DC Power & Load Modules
N6700C-N6705C Mainframes
30 Specifications Guide
Arbitrary Waveform Generator Maximum Bandwidth
NOTE
The information in this section only applies when the power modules are
installed in a Keysight N6705 DC Power Analyzer.
The following tables characterize the maximum bandwidth of the
arbitrary waveform generator. The maximum bandwidth is based on a
sine wave into a resistive load and apples to any output current. The
following definitions apply in the frequency tables:
V p-p =
Voltage peak-to-peak
3 dB max. =
Max. frequency where the voltage drops to 3 dB below its setting
THD 3 dB =
The total harmonic distortion at 3 dB max. frequency
THD < 1.5% =
The frequency below which the THD is less than 1.5%.
Voltage
3 dB max
THD 3 dB
THD < 1.5%
N6751A & N6752A
0.5 Vp-p 4000 Hz 12% 440 Hz
1.0 Vp-p 2200 Hz 21% 440 Hz
2.5 Vp-p 900 Hz 25% 265 Hz
5.0 Vp-p 500 Hz 27% 160 Hz
50.0 Vp-p 340 Hz 22% 25 Hz
N6753A & N6755A
0.2 Vp-p 2300 Hz 10% 1300 Hz
0.4 Vp-p 1500 Hz 15% 800 Hz
1.0 Vp-p 980 Hz 19% 480 Hz
2.0 Vp-p 580 Hz 21% 300 Hz
20.0 Vp-p 400 Hz 12% 32 Hz
N6754A & N6756A
0.6 Vp-p 2800 Hz 8.0% 1600 Hz
1.2 Vp-p 1400 Hz 15% 800 Hz
3.0 Vp-p 600 Hz 17% 300 Hz
6.0 Vp-p 400 Hz 20% 200 Hz
60.0 Vp-p 344 Hz 12% 30 Hz