Datasheet
N6731B-N6792A DC Power & Load Modules
N6700C-N6705C Mainframes
60 Specifications Guide
Supplemental Characteristics (continued)
N6791A
N6792A
Protection Accuracy:
Programmable over-current
High range
Low range
0.2 % +0.1A
0.2 % +0.1A
0.2 % +0.1A
0.2 % +0.1A
Fixed over-current
High range
Low range
21.42 A ±0.15 A
2.142 A ±0.15 A
42.84 A ±0.24 A
4.284 A ±0.20 A
Fixed over-voltage 66 V ±0.35 V 66 V ±0.35 V
Fixed over-power 110 W 220 W
N6705 Oscilloscope Measurement Accuracy
(@ 25 °C ±5 °C. Applies to any individual point of the trace.)
Current high range
0.04 +16 mA 0.04 +23 mA
Current low range
0.04 +14 mA 0.04 +20 mA
Voltage
0.03 +13 mV 0.03 +13 mV
Measurement Bandwidth:
Voltage & Current (rms)
-3 dB
-1 dB
20 kHz
10 kHz
20 kHz
10 kHz
Output Ripple and Noise:
Current (rms)
2 mA 3 mA
Voltage (rms) 5 mV 5 mV
Programmable Short/Open:
Programmable short 80 mΩ 40 mΩ
Input Off Impedance 500 kΩ 500 kΩ
Reverse Current Capability:
With the input off, load modules can tolerate reverse currents up the rated
current input of the module.
Series and Parallel Operation:
Identically rated load modules can be operated directly in parallel. Load modules
cannot
be used in series with other load modules or any other N67xx modules.
Output Quadrant Characteristic