Datasheet

N6731B-N6792A DC Power & Load Modules
N6700C-N6705C Mainframes
58 Specifications Guide
Performance Specifications
N6791A
N6792A
Input Ratings:
Current 0 - 20 A 0 - 40 A
Voltage 0 60 V 0 - 60 V
Maximum Power @ 40 °C
NOTE 1
100 W 200 W
Specified Current @ Low Voltage Operation:
(see Output Quadrant Characteristic)
1.6 V 20 A 40 A
1 V 12.5 A 25 A
0.5 V 6.25 A 12.5 A
0.1 V 1.25 A 2.5 A
Programming Accuracy:
(@ 25 °C ±5 °C after a 30 minute warm-up.
Applies from minimum to maximum programming range at any load.)
Current high range 20 A/40 A 0.04 % + 2.6 mA 0.04 % + 5.2 mA
Current low range 2 A/4 A 0.04 % + 0.46 mA 0.04 % + 0.92 mA
Voltage 60 V 0.03 % + 7.2 mV 0.03 % + 7.2 mV
Resistance high range
NOTE 2
8 kΩ ± (0.1 % + 0.0014) S ± (0.1 % + 0.0028) S
Resistance medium range
NOTE 2
100 Ω ± (0.1 % + 0.014) S ± (0.1 % + 0.019) S
Resistance low range
NOTE 2
3 Ω ± (0.1 % + 0.38) S ± (0.1 % + 0.55) S
Power high range 100 W/200 W 0.06 % + 180 mW 0.06 % + 360 mW
Power low range 10 W/20 W 0.06 % + 30 mW 0.06 % + 65 mW
Measurement Accuracy:
(@ 25 °C ±5 °C.
Applies when measuring the default value of 1024 data points with a 20.48 μs time interval.)
Current high range 20 A/40 A 0.04 % + 2.4 mA 0.04 % + 4.6 mA
Current low range 2 A/4 A 0.04 % + 0.40 mA 0.04 % + 0.82 mA
Voltage 60 V 0.03 % + 7.2 mV 0.03 % + 7.2 mV
Power high range 100 W/200 W 0.06 % + 160 mW 0.06 % + 320 mW
Power low range 10 W/20 W 0.06 % + 25 mW 0.06 % + 50 mW
1
Maximum continuous power is derated linearly from 100% of maximum @ 40 °C
to 75% of maximum @ 55 °C.
2
Based on minimum operating resistance, 0.08 Ω for 100W and 0.04 Ω for 200W. This means that below 1.6 V, the rated
current decreases linearly based on the rate of the respective minimum operating resistance.