Datasheet
N6731B-N6792A DC Power & Load Modules
N6700C-N6705C Mainframes
32 Specifications Guide
Performance Specifications
N6761A/N6762A
N6763A/N6765A
N6764A/N6766A
DC Ratings:
Voltage 0 - 50 V 0 - 20 V 0 - 60 V
Current
NOTE 1
0 - 1.5 A / 0 - 3 A 0 - 50 A 0 - 20 A / 0 - 17 A
Power 50 W / 100 W 300 W / 500 W 300 W / 500 W
Low programming ranges (V & I) 5.5 V; 100 mA - -
Low measurement ranges (V & I) 5.5 V; 100 mA 2 V; 1.5 A 6 V; 0.5 A
Output Ripple and Noise (PARD):
(from 20 Hz – 20 MHz)
CV peak-to-peak 4.5 mV 5 mV 6 mV
CV rms 0.35 mV 1 mV 1 mV
Load Effect (Regulation):
(Applies for any output load change, with a maximum load-lead drop of 1V/lead.
The load lead drop reduces the maximum available voltage at the load.
)
Voltage 0.5 mV 2 mV 2 mV
Current
30 µA (@ 0 –7 V)
65 µA (@ 7 – 50V)
12 mA 5 mA
Source Effect (Regulation):
Voltage 0.5 mV 0.5 mV 1.2 mV
Current
30 µA
5 mA 2 mA
Programming Accuracy:
(@ 23 °C ±5 °C after 30 minute warm-up.
Applies from minimum to maximum programming range at any load.)
Voltage, high range 0.016% + 6 mV 0.03% + 5 mV 0.03% + 12 mV
Voltage, low range 0.016% + 1.5 mV - -
Current, high range
0.04% + 200 µA
0.1% + 15 mA
0.075% + 4 mA
Current, low range
0.04% + 30 µA (@ 0 –7 V)
0.04% + 55 µA (@ 7 –50 V)
- -
Voltmeter/Ammeter Measurement Accuracy:
(@ 23 °C ±5 °C)
Applies when measuring 4096 data points with a 20.48
μs time interval.)
Voltage, high range 0.016% + 6 mV 0.03% + 10 mV 0.03% + 25 mV
Voltage, low range 0.016% + 1.5 mV 0.03% + 1.5 mV 0.03% + 5 mV
Current, high range
0.04% + 160 µA
0.1% + 10 mA 0.1% + 5 mA
Current, low range
0.03% + 15
µ
A (@ 0 – 7 V)
0.03% + 55 µA (@ 7 – 50 V)
0.05% + 1.1 mA
NOTE 2
0.05% + 0.75 mA
NOTE 2
200 µA current range (Option 2UA) 0.5% + 100 nA - -
Load Transient Recovery Time:
(time to recover to within the settling band following a load change
- from 60% to 100% and from 100% to 60% of full load for model N6761A
- from 50% to 100% and from 100% to 50% of full load for model N6762A through N6766A)
Settling band
± 75 mV
± 30 mV
NOTE 3
± 90 mV
NOTE 4
Time
< 100 µs
< 100 µs
< 100 µs
1
Output current is derated 1% per °C above 40°C.
2
Applies when measuring currents that remain within the low range. Due to thermal settling, when transitioning
from measuring full-rated output current (the worst case), to measuring the current within the low range, the low
range accuracy specification is typically met within 5 seconds after the current has transitioned into the low
range. Accuracies within this 5 second settling period are typically 2X the specified accuracy or better.
3
When relay option 760 or 761 is installed on Models N6763Aand N6765A, the settling band is ±200 mV.
4
When relay option 760 or 761 is installed on Models N6764Aand N6766A, the settling band is ±350 mV.