Probe characteristics
Bandwidth (–3 dB) Zoom-out channel: DC to 3 MHz
Zoom-in channel: DC to 500 kHz
Risetime (Tr = 0.35/bandwidth, 10 to 90%) Zoom-out channel: < 0.116 µsec
Zoom-in channel: < 0.7 µsec
Minimum measurable current
1
250 μA (with N2822A 20 m, 500 mW)
50 μA (with N2824A 100 mΩ, 500 mW)
5 mA (with N2825A user-defined 1 mΩ, 500 mW)
500 nA (with N2825A user-defined 1 k, 500 mW)
Maximum measurable current 5 A (with N2822A 20 mΩ, 500 mW)
2.2 A (with N2824A 100 mΩ, 500 mW)
5 A
2
(with N2825A user-defined 1 mΩ, 500 mW)
1.2 mA
2
(with N2825A user-defined 1 k, 500 mW)
Gain
5
Zoom-in channel: 300 ± 3%
Zoom-out channel: 1.97 ± 3%
Output voltage rate Zoom-in channel: 6 V/A (with N2822A 20 mΩ)
30 V/A (with N2824A 100 mΩ)
Zoom-out channel: 0.04 V/A (with N2822A 20 mΩ)
0.2 V/A (with N2824A 100 mΩ)
DC amplitude accuracy ± 3% or 10 μA (whichever is greater)
Max input voltage ± 12 V
Burden voltage (voltage drop on Rsense) Measured current
1
Rsense
Dynamic range 20,000:1, 86 dB (N2820A)
1,000:1, 60 dB (N2821A)
Output impedance 1 MΩ
Noise (ACrms) 150 μA (with N2822A 20 mΩ, hi res mode on)
3
240 μA (with N2822A 20 mΩ, hi res mode off)
30 μA (with N2824A 100 mΩ, hi res mode on)
3
50 μA (with N2824A 100 mΩ, hi res mode off)
Sensor resistor accuracy ± 1% (N2822A, N2824A)
Sensor resistor power rating 500 mW (N2822A, N2824A)
Temperature coefficient of sensor resistor < 20 ppm/°C
Cable lengths Sensor leads: 16 cm
Probe cable: 1.2 m
Temperature Operating: 0 to +40 °C
Non-operating: –40 to +70 °C
Humidity Operating: tested at 95% RH at +40 °C
Non-operating: tested at 90% RH, +65 °C
ESD 8 kV HBM
Standard accessories 1 each 20 mΩ resistor sensor head 5 each MBB headers
1 each 100 mΩ resistor sensor head 5 each MBB receptacles
1 each user defined resistor sensor head 1 each ground lead
5 each twisted leads (22 AWG) with sockets 1 each screw driver
5 each twisted leads (22 AWG) without sockets 1 each passive cable (with N2820A only)
1 each user guide manual (English)
Compatible oscilloscopes InfiniiVision 3000 X-Series (with software version 2.30 or higher)
InfiniiVision 4000 X-Series (with software version 3.10 or higher) and 6000 X-Series
Infiniium S, 9000A, 9000 H-Series (with software version 4.20 or higher)
4
1. Vsupply is equal to 5 V, solder attached.
2. Max current varies with max resistor power rating. The examples in the table assume 500 mW power rating.
3. With scope with high resolution mode on, sampling rate < 2.5 MSa/s, 20 MHz low pass filter on.
4. Infiniium V, 90000A, 90000 X-, and 90000 Q-Series are not compatible with N2820A/21A current probe.
5. Denotes warrantied specification after 20-minute warm up. All others entries in the table are characteristics.