Datasheet
03 | Keysight | N2795A/96A/97A Single-ended Active Probes - Data Sheet
N2795A N2796A N2797A
Probe bandwidth*
(–3 db)
1 GHz 2 GHz 1.5 GHz**
Risetime
(calculated,
10-90%)
350 psec 175 psec 233 psec
System
bandwidth
(with Keysight
oscilloscope)
500/600 MHz (with
Keysight’s 500/600
MHz InfiniiVision/
Infiniium
oscilloscope)
1 GHz (with
Keysight’s
1 GHz
InfiniiVision/
Infiniium
oscilloscope)
1 GHz (with Keysight’s
1 GHz InfiniiVision/
Infiniium
oscilloscope)
Attenuation ratio
(@DC)
10:1 ± 0.5%
Input dynamic
range
–8 V to +8 V (DC or peak AC)
Non-destructive
max input voltage
–20 V to +20 V
Offset range ±8 V ±12 V ±12 V
DC offset error
(Output zero)
< ±1 mV
Flatness
(at 25 °C)
Typical 0.4 dB
(100 kHz - 100 MHz)
Typical 0.6 dB
(100 MHz - 500 MHz)
Typical 0.8 dB
(500 MHz - 1 GHz)
Typical 1.3 dB
(1 GHz - 2 GHz)
Typical 0.3 dB
(10 Hz – 100 MHz)
Typical 0.8 dB
(100 MHz – 1 GHz)
Typical 2.0 dB
(1 GHz – 1.5 GHz)
Flatness over
temperature
(–40 to +85 °C)
Typical 0.3 dB
(10 Hz – 100 kHz)
Typical 0.6 dB
(100 kHz – 100 MHz)
Typical 0.8 dB
(100 MHz – 500 MHz)
Typical 2.0 dB
(500 MHz – 1 GHz)
Typical 2.5 dB
(1 GHz – 1.5 GHz)
Input resistance* 1 MΩ +0 %, –2.5 % 1 MΩ ±3%
Input
capacitance
1 pF
Probe noise < 2.5 mVrms (referred to input)
Output
impedance
50 Ω
N2795A N2796A N2797A
Internal power Keysight Autoprobe Interface from scope
(InfiniiVision and Infiniium)
Cable length 1.3 m 2 m
Probe weight Approx. 100 g Approx. 108 g
Ambient
operating
Temperature
0 to 50 °C –40 to +85 °C
Ambient
non-operating
temperature
–40 to 70 °C –40 to +85 °C
Operating
humidity
95% RH @ 40 °C
Non-operating
humidity
90% RH @ 65 °C
Operating altitude 4000 m
ESD 8 kV HBM
Standard
accessories
– 2 each spring probe tip
– 2 rigid probe tip
– 1 each flex nose clip adapter
(red and black)
– 1 each copper pad, 10x
– 1 each dual lead socketed adapter,
9 cm
–
1 each right angle ground, 5 cm
– 1 each right angle ground, 10 cm
– 2 each ground blade
– 1 each offset ground
– 1 each flex ground
– 4 color coded rings (each yellow,
green, blue and purple)
– 10 each solderable tip
– 5 each rigid probe tip
– 2 each flex nose clip
adapter (red and black)
– 1 each pico hook tip (red
and black)
– 2 each dual lead socket-
ed adapter,
9 cm (800 MHz)
– 2 each dual lead
socketed adapter,
6 cm (1 GHz)
– 2 each right angle
ground, 5 cm
– 2 each ground blade
– 4 color coded rings
(each yellow, green,
blue and purple)
Others (included) -1 each accessory configuration card
Compatible
Keysight scopes
Keysight InfiniiVision 3000 X-, 4000
X-
, 6000 X-, 5000, 6000, 7000 (except
6000 100MHz) and Infiniium S-Series,
9000, 90000A, 90000 X-, 90000 Q-,
Z-Series(with N5442A)
Keysight InfiniiVision
3000 X-, 4000 X-,
6000 X-, and Infiniium
S-Series, 9000, 90000A,
90000 X-, 90000 Q-,
Z-Series (with N5442A)
*
denotes warranted electrical specifications after 20 minute warm-up, all others are typical
** typical 2 GHz, when used with rigid probe tip, ground blade and handheld
Use ex nose clip adapters with the dual lead adapter to obtain access to IC leads
or head connectors.
The dual lead adapter allows you to easily connect the probe to a popular 0.1” pin
header with 0.025” square pins.