Datasheet

03 | Keysight | N2795A/96A/97A Single-ended Active Probes - Data Sheet
N2795A N2796A N2797A
Probe bandwidth*
(–3 db)
1 GHz 2 GHz 1.5 GHz**
Risetime
(calculated,
10-90%)
350 psec 175 psec 233 psec
System
bandwidth
(with Keysight
oscilloscope)
500/600 MHz (with
Keysight’s 500/600
MHz InfiniiVision/
Infiniium
oscilloscope)
1 GHz (with
Keysight’s
1 GHz
InfiniiVision/
Infiniium
oscilloscope)
1 GHz (with Keysight’s
1 GHz InfiniiVision/
Infiniium
oscilloscope)
Attenuation ratio
(@DC)
10:1 ± 0.5%
Input dynamic
range
–8 V to +8 V (DC or peak AC)
Non-destructive
max input voltage
–20 V to +20 V
Offset range ±8 V ±12 V ±12 V
DC offset error
(Output zero)
< ±1 mV
Flatness
(at 25 °C)
Typical 0.4 dB
(100 kHz - 100 MHz)
Typical 0.6 dB
(100 MHz - 500 MHz)
Typical 0.8 dB
(500 MHz - 1 GHz)
Typical 1.3 dB
(1 GHz - 2 GHz)
Typical 0.3 dB
(10 Hz – 100 MHz)
Typical 0.8 dB
(100 MHz – 1 GHz)
Typical 2.0 dB
(1 GHz – 1.5 GHz)
Flatness over
temperature
(–40 to +85 °C)
Typical 0.3 dB
(10 Hz – 100 kHz)
Typical 0.6 dB
(100 kHz – 100 MHz)
Typical 0.8 dB
(100 MHz – 500 MHz)
Typical 2.0 dB
(500 MHz – 1 GHz)
Typical 2.5 dB
(1 GHz – 1.5 GHz)
Input resistance* 1 MΩ +0 %, –2.5 % 1 MΩ ±3%
Input
capacitance
1 pF
Probe noise < 2.5 mVrms (referred to input)
Output
impedance
50 Ω
N2795A N2796A N2797A
Internal power Keysight Autoprobe Interface from scope
(InfiniiVision and Infiniium)
Cable length 1.3 m 2 m
Probe weight Approx. 100 g Approx. 108 g
Ambient
operating
Temperature
0 to 50 °C –40 to +85 °C
Ambient
non-operating
temperature
–40 to 70 °C –40 to +85 °C
Operating
humidity
95% RH @ 40 °C
Non-operating
humidity
90% RH @ 65 °C
Operating altitude 4000 m
ESD 8 kV HBM
Standard
accessories
– 2 each spring probe tip
– 2 rigid probe tip
1 each flex nose clip adapter
(red and black)
– 1 each copper pad, 10x
1 each dual lead socketed adapter,
9 cm
1 each right angle ground, 5 cm
– 1 each right angle ground, 10 cm
– 2 each ground blade
– 1 each offset ground
– 1 each flex ground
4 color coded rings (each yellow,
green, blue and purple)
– 10 each solderable tip
– 5 each rigid probe tip
2 each flex nose clip
adapter (red and black)
1 each pico hook tip (red
and black)
2 each dual lead socket-
ed adapter,
9 cm (800 MHz)
2 each dual lead
socketed adapter,
6 cm (1 GHz)
2 each right angle
ground, 5 cm
2 each ground blade
4 color coded rings
(each yellow, green,
blue and purple)
Others (included) -1 each accessory configuration card
Compatible
Keysight scopes
Keysight InfiniiVision 3000 X-, 4000
X-
, 6000 X-, 5000, 6000, 7000 (except
6000 100MHz) and Infiniium S-Series,
9000, 90000A, 90000 X-, 90000 Q-,
Z-Series(with N5442A)
Keysight InfiniiVision
3000 X-, 4000 X-,
6000 X-, and Infiniium
S-Series, 9000, 90000A,
90000 X-, 90000 Q-,
Z-Series (with N5442A)
*
denotes warranted electrical specifications after 20 minute warm-up, all others are typical
** typical 2 GHz, when used with rigid probe tip, ground blade and handheld
Use ex nose clip adapters with the dual lead adapter to obtain access to IC leads
or head connectors.
The dual lead adapter allows you to easily connect the probe to a popular 0.1” pin
header with 0.025” square pins.