Datasheet
Page 9
Find us at www.keysight.com
Characteristics, non-condensing and specications
Model number N2750A N2751A N2752A
Differential mode rejection ratio 40 dB at 1 MHz
40 dB at 30 MHz
30 dB at 1 GHz
Channel isolation, at single-ended
mode
60 dB at 1 MHz
45 dB at 30 MHz
20 dB at 1 GHz
Offset gain accuracy < 3 % before calibration
< 1 % after calibration
Noise referred to input 2.5 mVrms (@2:1), 7 mVrms (@10:1)
Maximum signal slew rate 30 V/ns (differential, 10x),
15 V/ns (single-ended, 10x)
Zero offset error referred to input < 30 mV before calibration scope,
< 5 mV after calibration scope
Propagation delay 6.85 nsec
ESD tolerance > 8 kV from 100 pF, 300 Ω, HBM
Output impedance 50 Ω
Cable length 1.3 m
Probe weight 100 g
Ambient operating temperature 0 to +40 ºC
Ambient non-operating temperature –40 to +70 ºC
Operating humidity Up to 95 % RH @ +40 ºC, non-condensing
Non-operating humidity Up to 90 % RH @ +65 ºC, non-condensing
Operating altitude 4,600 m
Non-operating altitude 15,300 m
EMC/EMI Radiated Emission: CISPER 11 Group 1, Class A.
ESD: IEC61000-4-2 4KV contact discharge, 8KV air discharge
Standard accessories 2x InniiMode solder-in tips
1x InniiMode socketed tip
1x Differential browser tip
Others (included) 1x calibration certicate
1x accessory conguration card
Recommended Keysight
oscilloscopes***
InniiVision 3000 X-Series with
software ver 2.2 or higher,
4000 X-Series with 3.01 or higher,
6000 X-Series with 6.0 or higher,
Inniium 9000 Series with 3.5 or
higher, S-Series with 5.0 or higher
Inniium 9000 and 90000A Series
with software ver 3.5 or higher,
S-Series with 5.0 or higher
Inniium 9000 and 90000A Series
with software ver 3.5 or higher,
S-Series with 5.0 or higher
* Denotes warranted electrical specications at 2:1 attenuation mode after 20 minute warm-up, All others are typical.
** When calibrated on the oscilloscope, these characteristics are determined by the oscilloscope characteristics.
*** The N2750A Series differential probes can be used with the Inniium 90000X/Q Series scope and the N5442A precision BNC adapter.
**** Assumes symmetric differential signals.