Dual-channel WaveGen — built-in function/arbitrary waveform generator (Specification are typical) (Continued)
Amplitude Range: Minimum 20 mVpp if |offset| ≤ 0.5 Vpp into Hi-Z
1
10 mVpp if |offset| ≤ 0.5 Vpp into 50 Ω
1
Range: Maximum 10 Vpp except, 9 Vpp if sinc or cardiac, 7.5 Vpp if Gaussian pulse into Hi-Z;
5 Vpp/4.5 Vpp into 50 Ω
Resolution 100 μV or 3 digits, whichever is higher
Accuracy 1.5% (frequency = 1 kHz)
DC offset Range ± 5 V into Hi-Z, except ± 4 V if sine wave, ± 2.5 V if sinc, cardiac, or Gaussian
pulse into Hi-Z
± 2.5 V into Hi-Z, except ± 2 V if sine wave, ± 1.25 V if sinc, cardiac, or
Gaussian pulse into 50 Ω
Resolution Larger of 250 uV or 3 digits
Accuracy (waveform modes) ± 1.5% of offset setting ± 1% of amplitude ± 1 mV
Accuracy (DC mode) ± 1.5% of offset setting ± 3 mV
Main output Impedance 50 Ω typical
Isolation Not available, main output BNC is grounded
Protection Overload automatically disables output
Trigger output Trigger output available on trig-out BNC
Quick action customization key
Quick measure all Displays a popup containing a snapshot of all the single waveform measurements
Quick measure statistics reset Resets all measurement statistics and the measurement count
Quick mask statistics reset Resets mask statistics and counters
Quick histogram statistics reset Zeros the histogram counters
Quick print Print the current screen image
Quick save Saves the current setup, screen image, or data file as specified in the settings menu
Quick email Emails the current setup, screen image, or data file as specified in the settings menu
Quick recall Recalls setup, mask, or reference waveform
Quick freeze display Freezes the display without stopping running acquisitions or unfreezes the display if currently frozen.
Waveform intensity preserved
Quick trigger mode Toggles the trigger mode between auto and normal
Quick clear display Clears the display
Display characteristics
Display 12.1-inch capacitive multi-touch/gesture enabled color TFT LCD
Display mode Zone/zoom/annotation mode and waveform placement mode
Resolution 800 (H) x 600 (V) pixel format (screen area)
Graticules 8 vertical divisions by 10 horizontal divisions with intensity controls
Format YT and XY
Maximum waveform update rate > 135,000 wfm/s (real time)
> 450,000 wfm/s (real time max update rate)
Persistence Off, infinite, variable persistence (100 ms to 60 s)
Intensity gradation 256 intensity levels
1. Sinc, cardiac and Gaussian pulse: ± 1.25 V into Hi-Z; ± 625 mV into 50 Ω.