Search, navigate, and lister
Type Edge, pulse width, rise/fall, runt, frequency peak, serial bus 1, serial bus 2
Copy Copy to trigger, copy from trigger
Frequency peak Source Math functions
Max number of peaks 11
Control Threshold, excursion, results order (frequency or amplitude)
Result display Event lister or navigation. Manual or autoscroll via navigation or touch event lister entry to jump to a
specific event
Waveform measurements
DC vertical accuracy/cursors
2
Single cursor accuracy: ± [DC vertical gain accuracy + DC vertical offset accuracy + 0.21% full scale]
Dual cursor accuracy: ± [DC vertical gain accuracy + 0.42% full scale]
1
Number of measurements 56 automatic measurements, maximum of 10 displayed at a time
Cursors 2 pairs of XY cursors
Automatic measurement of positions, ΔX, 1/ΔX, ΔY, and ΔY/ΔX
Automatic
measurements
Measurements continuously updated with statistics. Cursors track last selected measurement.
Select up to 10 measurements from the list below:
Snapshot Makes a snapshot of 31 most popular measurements. Touchable target to populate the
measurement side bar
Voltage/current Peak-to-peak, maximum, minimum, amplitude, top, base, overshoot, preshoot, average- N cycles,
average-full screen, DC RMS-N cycles, DC RMS-full screen, AC RMS-N cycles, AC RMS-full screen
(standard deviation), ratio-N cycles, ratio-full screen
Time Period, frequency, counter, + width, - width, burst width, + duty cycle, - duty cycle, bit rate, rise
time, fall time, delay, phase, X at min Y, X at max Y
Count Positive pulse count, negative pulse count, rising edge count, falling edge count
Mixed Area-N cycles, area-full screen
Jitter Option DSOX6JITTER required
Data TIE, clock TIE, N-period, period-period, + width to + width, - width to - width
Real-time eye Option DSOX6JITTER required
Eye width, eye height
Dual-channel (requires
N2820A probe)
Charge-N cycles, charge-full screen, peak-peak, amplitude, DC RMS-N cycles, DC RMS-full screen,
AC RMS- N cycles, AC RMS-full screen (standard deviation), average-N cycles, average-full screen,
base
Automatic
measurement logging
Available via BV0004B BenchVue
Counter Built-in frequency counter (see “Precision counter/totalizer section” for the 10-digit counter)
Source Any analog and digital channel
Resolution 5 digits
Max frequency 1 GHz (1.2 GHz typical)
1. Denotes warranted specifications; All others are typical. Specifications are valid after a 30-minute warm-up period and ± 10 °C from firmware calibration
temperature.
2. 1 mV/div is a magnification of 2 mV/div setting. For vertical accuracy calculations, use full scale of 16 mV for 1 mV/div sensitivity setting.
Performance Characteristics (Continued)