Waveform math
Operators Add, subtract, multiply, FFT
FFT Windows: Hanning, flat top, rectangular; Blackman-Harris - up to 64 kpts resolution
Sources Math functions available between any two channels
Display characteristics
Display 8.5-inch WVGA
Resolution 800 (H) x 480 (V) pixel format (screen area)
Interpolation Sin(x)/x interpolation (using FIR filter; used when there is less than one sample per column of the display)
Persistence Off, infinite, variable persistence (100ms-60s)
Intensity gradation 64 intensity levels
Modes Normal
XY - XY mode changes the display from voltage versus time scale to a volts versus volts scale
Roll – Displays the waveform moving across the screen from right to left much like a strip chart recorder
MSO (digital channels)
Upgradable from DSO Yes
MSO channels 8 channels (D0 to D7)
Maximum sample rate 1 GSa/s
Maximum record length 50kpts per channel (digital channels only)
12.5kpts per channel (analog and digital channels)
Threshold selections TTL (+1.4 V), CMOS (+2.5 V), ECL (-1.3 V), User-definable (±8.0 V in 10 mV stops)
Threshold accuracy* ± (100 mV + 3% of threshold settings)
Maximum input dynamic
range
±10 V about threshold
Minimum voltage swing 500 mVpp
Input impedance 100 kΩ ± 2% at probe tip, ~8 pF
Minimum detectable pulse
width
5ns
Channel-to-channel skew 2 ns (typical), 3 ns (maximum)
Environmental and safety
Power line consumption 100 watts
Temperature Operating: 0 to +55 °C
Non-operating: -30 to +71 °C
Humidity Operating: Up to 80% RH at or below +40 °C; up to 45% RH up to +50 °C
Non-operating: Up to 95% RH up to 40 °C; up to 45% RH up to 50 °C
Altitude Operating and non-operating: up to 4,000 m
Electromagnetic
compatibility
Meets EMC Directive (2004/108/EC), meets or exceeds IEC 61326-1:2005/EN
61326-1:2006 Group 1 Class A requirement
CISPR 11/EN 55011
IEC 61000-4-2/EN 61000-4-2
IEC 61000-4-3/EN 61000-4-3
IEC 61000-4-4/EN 61000-4-4
IEC 61000-4-5/EN 61000-4-5
IEC 61000-4-6/EN 61000-4-6
IEC 61000-4-11/EN 61000-4-11
Canada: ICES-001:2004
Australia/New Zealand: AS/NZS
Safety UL61010-1 2nd edition, CAN/CSA22.2 No. 61010-1-04
Oscilloscopes redefined: Breakthrough technology delivers more scope for the same budget
Performance characterisitics
* Denotes warranted specifications, all others are typical.
Specifications are valid after a 30-minute warm-up period and from ±10 °C firmware calibration temperature
.