Datasheet
20 | Keysight | InfiniiVision 2000 X-Series Oscilloscopes - Data Sheet
Connectivity
Standard ports One USB 2.0 hi-speed device port on rear panel. Supports USBTMC protocol
Two USB 2.0 hi-speed host ports, front and rear panel
Supports memory devices, printers and keyboards
Optional ports GPIB, LAN, WVGA video out
General and environmental characteristics
Power line consumption 100 W
Power voltage range 100 to 120 V, 50/60/400 Hz; 100 to 240 V, 50/60 Hz ± 10% auto ranging
Temperature Operating: 0 to +55 °C
Non-operating: –30 to +71 °C
Humidity Operating: Up to 80% RH at or below +40 °C; up to 45% RH up to +50 °C
Non-operating: Up to 95% RH up to 40 °C; up to 45% RH up to 50 °C
Altitude Operating: Up to 4,000 m, Non-operating 15,300 m
Electromagnetic compatibility Meets EMC Directive (2004/108/EC), meets or exceeds IEC 61326-1:2005/EN
61326-1:2006 Group 1 Class A requirement
CISPR 11/EN 55011
IEC 61000-4-2/EN 61000-4-2
IEC 61000-4-3/EN 61000-4-3
IEC 61000-4-4/EN 61000-4-4
IEC 61000-4-5/EN 61000-4-5
IEC 61000-4-6/EN 61000-4-6
IEC 61000-4-11/EN 61000-4-11
Canada: ICES-001:2004
Australia/New Zealand: AS/NZS
Safety UL61010-1 2nd edition, CAN/CSA22.2 No. 61010-1-04
Dimensions (W x H x D) 381 mm (15 in) x 204 mm (8 in) x 142 mm (5.6 in)
Weight Net: 3.9 kg (8.5 lbs), shipping: 4.1 kg (9.0 lbs)
Nonvolatile storage
Reference waveform display 2 internal waveforms or USB thumb drive
Waveform storage Set up, .bmp, .png, .csv, ASCII, XY, reference waveforms, .alb, .bin, lister, mask, HDFS
Max USB flash drive size Supports industry standard flash drives
Set ups without USB flash drive 10 internal setups
Set ups with USB flash drive Limited by size of USB drive
InfiniiVision X-Series Physical Characteristics
Included standard with oscilloscope
Standard secure erase
Standard probe
– N2862B 150 MHz 10:1 passive probe Standard one per channel for 70 and 100 MHz models
– N2863B 300 MHz, 10:1 passive probe Standard one per channel for 200 MHz models
– N6459-60001 8-channel logic probe and accessory kit Standard on MSO models or with DSOX2MSO upgrade
Built-in help language support for English, Japanese, simplified Chinese, traditional Chinese, Korean, German, French, Spanish, Russian, Portuguese and
Italian, Certificate of Calibration, Documentation CD
Interface language support GUI menus: English, Japanese, simplified Chinese, traditional Chinese, Korean, German, French, Spanish, Russian,
Portuguese, Thai, Polish and Italian
Localized power cord
For MET/CAL procedures, click on the Cal Labs solutions link below Cal Labs Solutions http://www.callabsolutions.com/products/Keysight/.
These procedures are FREE to customers.