Datasheet
15 | Keysight | InfiniiVision 2000 X-Series Oscilloscopes - Data Sheet
Performance Characteristics
Specification overview
2002A 2004A 2012A 2014A 2022A 2024A
Bandwidth
1
(–3 dB) 70 MHz 100 MHz 200 MHz
Calculated rise time (10 to 90%) ≤ 5 ns ≤ 3.5 ns ≤ 1.75 ns
Input channels DSOX 2 4 2 4 2 4
MSOX 2 + 8 4 + 8 2 + 8 4 + 8 2 + 8 4 + 8
Maximum sample rate
1
2 GSa/s half-channel interleaved, 1 GSa/s per channel
Maximum memory depth 1 Mpt per channel (standard)
Display size and type 8.5-inch WVGA with 64 levels of intensity grading
Waveform update rate 200,000 waveforms per second
Vertical system analog channels
Input coupling AC, DC
Input sensitivity range 1 mV/div to 5 V/div
2
Input impedance 1 MΩ ± 2% (11 pF)
Vertical resolution 8 bits (measurement resolution is 12 bits with averaging)
Dynamic range ± 8 divisions from center screen
Maximum input voltage 300 Vrms, 400 Vpk; transient overvoltage 1.6 kVpk
With N2862B or N2863B 10:1 probe: 300 Vrms
Frequency de-rating (assumes sine wave input): 400 Vpk until 40 kHz. Then de-rates at 20 db/
dec until 6 Vpk
DC vertical accuracy ± [DC vertical gain accuracy + DC vertical offset accuracy + 0.25% full scale]
2
DC vertical gain accuracy
1
± 3% full scale (≥ 10 mV/div); ± 4% full scale (< 10 mV/div)
2
DC vertical offset accuracy ± 0.1 div ± 2mV ± 1% of offset setting
Channel-to-channel isolation ≥ 40 dB from DC to maximum specified bandwidth of each model
Position/offset range 1 MΩ 1 mV to 200 mV/div: ± 2 V, > 200 mV to 5 V/div: ± 50 V
Hardware bandwidth limits Approximately 20 MHz (selectable)
Horizontal system analog channels
2002A 2004A 2012A 2014A 2022A 2024A
Time base range 5 ns/div to 50 s/div 2 ns/div to 50 s/div
Time base accuracy
1
25 ppm ± 5 ppm per year (aging)
Time base delay time range Pre-trigger Greater of 1 screen width or 200 μs (400 μs in interleaving mode)
Post-trigger 1 s to 500 s
Channel-to-channel deskew range ± 100 ns
Δ Time accuracy (using cursors) ± (time base accuracy
1
reading) ± (0.0016
1
screen width) ± 100 ps
1. Denotes warranted specifications, all others are typical. Specifications are valid after a 30-minute warm-up period and from ± 10 °C firmware calibration
temperature.
2. 1 mV/div and 2 mV/div is a magnification of 4 mV/div setting. For vertical accuracy calculations, use full scale of 32 mV for 1 mV/div and 2 mV/div sensitivity
setting.