Datasheet

Page 7
Find us at www.keysight.com
Specifications
Performance specifications E36311A E36312A E36313A
Power output 80 W 80 W 160 W
DC output
rating (0 to 40ºC)
1 2 3 1 2 3 1 2 3
0 to 6 V 0 to +25 V 0 to -25 V 0 to 6 V 0 to 25 V 0 to 25 V 0 to 6 V 0 to 25 V 0 to 25 V
0 to 5 A 0 to 1 A 0 to 1 A 0 to 5 A 0 to 1 A 0 to 1 A 0 to 10 A 0 to 2 A 0 to 2 A
Series mode voltage NA 50 V 50 V
Parallel mode current NA 2A 4A
Load regulation ± (% of output + offset)
Voltage < 0.01% +2 mV < 0.01% +2 mV < 0.01% +4 mV
Current < 0.01% +250 uA < 0.01% +250 uA < 0.01% +500 uA
Line regulation ±(% of output + offset)
Voltage < 0.01% +1 mV < 0.01% +1 mV < 0.01% +1 mV
Current < 0.01% +250 uA < 0.01% +250 uA < 0.01% +500 uA
Output ripple and noise (20 Hz to 20 MHz)
Normal mode voltage < 350 uVrms/2 mVpp < 350 uVrms/2 mVpp
< 350 uVrms/
2 mVpp
< 1 mVrms/ 5 mVpp
Accuracy 12 months (23 ºC ± 5 ºC)
Programming accuracy ±(% of output + offset)
Voltage 0.1% +5 mV 0.05% +20 mV 0.03%
+2 mV
0.03% +5 mV 0.03%
+3 mV
0.03% +5 mV
Current
0.1% + 10 mA
0.1% +4 mA 0.04%
+3 mA
0.04% +2 mA 0.05%
+4 mA
0.04% +3 mA
Readback accuracy ±(% of output + offset)
Voltage 0.1% +5 mV 0.05% +10 mV 0.04%
+2mV
0.04% +5 mV 0.04%
+3 mV
0.03% +5 mV
Current 0.1% +10 mA 0.1% +4 mA 0.04%
+3 mA
0.04% +3 mA 0.05%
+5 mA
0.04% +3 mA
Low range current
1
NA 0.25% +80 uA 0.25% +80 uA
Load transient recovery time
(Time to recover to within the settling band following a load change from 50% to 100% and from 100% to 50% of full load)
Voltage settling band 15 mV 15 mV 15 mV 30 mV 15 mV
Voltage settling band
(parallel mode)
NA 30 mV 30 mV
Time < 50 uS
1. 20mA for CH1, 10mA for CH2 and CH3