Datasheet
04 | Keysight | E36100B Series Programmable DC Power Supplies - Data Sheet
Tolerance % E36102B E36103B E36104B E36105B E36106B
DC output rating (0 to 40 °C)
Max. voltage 6 V 20 V 35 V 60 V 100 V
Max. current 5 A 2 A 1 A 0.6 A 0.4 A
Load regulation ± (% of output + offset)
Voltage <0.01% + 2 mV 3 mV 6 mV 10 mV 20 mV
Current <0.02% + 250 µA 100 µA 50 µA 30 µA 20 µA
Line regulation ± (% of output + offset)
Voltage <0.01% + 1 mV 2 mV 4 mV 7 mV 12 mV
Current <0.02% + 250 µA 100 µA 50 µA 30 µA 20 µA
Output ripple and noise (20 Hz to 20 MHz)
Voltage RMS 350 µV 0.8 mV 1.2 mV 1.55 mV 2.5 mV
Pk-Pk 8 mV 15 mV 20 mV 30 mV 40 mV
Accuracy 12 months (23 °C ± 5 °C)
Programming accuracy ± (% of output + offset)
Voltage 0.05% + 3 mV 8 mV 12 mV 20 mV 40 mV
Current 0.05% + 5 mA 1 mA 0.6 mA 0.4 mA 0.3 mA
Readback accuracy ± (% of output + offset)
Voltage 0.05% + 3 mV 5 mV 8 mV 12 mV 20 mV
Current 0.05% + 4 mA 1 mA 0.5 mA 0.3 mA 0.2 mA
Low range current 0.25% + 40 µA
(0-20 mA)
40 µA
(0-8 mA)
40 µA
(0-4 mA)
40 µA
(0-3 mA)
40 µA
(0-2 mA)
Load transient recovery time
(Time to recover to within the settling band following a load change from 50% to 100% and from 100% to 50% of full load)
Voltage settling band 15 mV 50 mV 87.5 mV 150 mV 250 mV
Time <50 μs <50 μs <50 μs <50 μs <50 μs
Performance Specifications