Datasheet

Page 9
Find us at www.keysight.com
The 3000T X-Series has excellent signal integrity, including full
bandwidth to 1 mV/div and the ability to get up to 12-bits of
resolution using the high resolution acquisition mode�
Some oscilloscopes in this class limit their bandwidth at smaller
volt-per-division settings without on-display user notifications�
This is likely to keep the noise acceptable at lower volt-per-
division settings�
Table 2 shows a comparison of the typical noise floor at 20 μ/div
between the normal and high resolution mode� You will notice
that the noise floor performance improves as much as five times
Figure 12� High resolution mode allows you to lower your noise and
increase your resolution up to 12-bits
Table 2. Noise comparison between the normal and high-resolution mode at 20 μ/div.
50 Ω 1 GHz bandwidth Vrms measurement (units = mV)
Vertical setting Normal mode High resolution mode Notes
1 mV 0.277 0.072 Some other manufacturers will limit their bandwidth significantly at these
vertical settings, but the Keysight 3000T X-Series provides full bandwidth
at all settings.
2 mV 0.277 0.072
5 mV 0.297 0.081
10 mV 0.352 0.081
20 mV 0.597 0.102
50 mV 1.500 0.340
100 mV 2.560 0.480
200 mV 5.500 1.050
500 mV 15.200 3.630
1 V 26.000 4.830
Discover: Excellent Signal Integrity Allows you to See More Signal Detail