Datasheet
Page 28
Find us at www.keysight.com
Performance Characteristics
Vertical system analog channels
Hardware bandwidth limits Approximately 20 MHz (selectable)
Input coupling AC, DC
Input impedance Selectable: 1 MΩ ± 1% (14 pF), 50 Ω ± 1.5%
Input sensitivity range 100 MHz ~ 500 MHz models: 1 mV/div to 5 V/div
2
(1 MΩ and 50 Ω)
1 GHz models: 1 mV/div to 5 V/div
2
(1 MΩ), 1 mV/div to 1 V/div (50 Ω)
Vertical resolution 8 bits (measurement resolution is 12 bits with averaging)
Maximum input voltage 135 Vrms; 190 Vpk
Probing technology allows testing of higher voltages. For example, the included N2843A 10:1 probe supports
testing up to 300Vrms
Use this instrument only for measurements within its specified measurement category (not rated for CAT II, III, IV).
No transient overvoltage allowed
DC vertical accuracy ± [DC vertical gain accuracy + DC vertical offset accuracy + 0.25% full scale]
2
DC vertical gain accuracy
1
± 2.0% full scale
2
DC vertical offset accuracy ± 0.1 div ± 2 mV ± 1% of offset setting
Channel-to-channel isolation > 100:1 from DC to maximum specified bandwidth of each model (measured with same V/div and coupling on
channels)
Offset range ± 2 V (1 mV/div to 200 mV/div)
± 50 V (> 200 mV/div to 5 V/div)
Vertical system digital channels
Digital input channels 16 digital (D0 to D15. pod 1: D7 ~ D0, Pod 2: D15 ~ D8)
Thresholds Threshold per pod
Threshold selections TTL (+1.4 V), 5 V CMOS (+2.5 V), ECL (–1.3 V), user-defined (selectable by pod)
User-defined threshold range ± 8.0 V in 10 mV steps
Maximum input voltage ± 40 V peak CAT I
Threshold accuracy
1
± (100 mV + 3% of threshold setting)
Maximum input dynamic range ± 10 V about threshold
Minimum voltage swing 500 mVpp
Input impedance 100 kΩ ± 2% at probe tip
Input capacitance ~8 pF
Vertical resolution 1 bit
1� Denotes warranted specifications, all others are typical�
2� Specifications are valid after a 30-minute warm-up period and ± 10 °C from firmware calibration temperature� 1 mV/div and 2 mV/div are a magnification of
4 mV/div setting� For vertical accuracy calculations, use full scale of 32 mV for 1 mV div and 2 mV/div sensitivity setting�
DSO and MSO 3000 X-Series oscilloscopes
3000T X-Series specification overview
3012T 3014T 3022T 3024T 3032T 3034T 3052T 3054T 3102T 3104T
Bandwidth
1
(–3 dB) 100 MHz 200 MHz 350 MHz 500 MHz 1 GHz
Calculated rise time (10 to 90%) ≤ 3.5 ns ≤ 1.75 ns ≤ 1 ns ≤ 700 ps ≤ 450 ps
Input channels
DSOX 2 4 2 4 2 4 2 4 2 4
MSOX 2 + 16 4 + 16 2 + 16 4 + 16 2 + 16 4 + 16 2 + 16 4 + 16 2 + 16 4 + 16
Maximum sample rate 5 GSa/s half channels, 2.5 GSa/s all channels
Maximum memory depth Standard 4 Mpts, Standard segmented memory
Display size and type 8.5-inch capacitive touch gesture-enabled display
Waveform update rate > 1,000,000 waveforms per second