Datasheet

Page 26
Find us at www.keysight.com
DSO and MSO 4000 X-Series oscilloscopes
4000 X-Series specification overview
4022A 4024A 4032A 4034A 4052A 4054A 4104A 4154A
Bandwidth
1
(–3 dB) 200 MHz 350 MHz 500 MHz 1 GHz 1.5 GHz
3
All-channel real-time bandwidth 200 MHz 350 MHz 500 MHz 1 GHz 1 GHz
Calculated rise time (10 to 90%) ≤ 1.75 ns ≤ 1 ns ≤ 700 ps ≤ 450 ps ≤ 300 ps
Input channels DSOX 2 4 2 4 2 4 4 4
MSOX 2 + 16 4 + 16 2 + 16 4 + 16 2 + 16 4 + 16 4 + 16 4 + 16
Maximum sample rate 5 GSa/s half channel, 2.5 GSa/s all channel
Maximum memory depth Standard 4 Mpts, standard segmented memory
Display size and type 12.1-inch high-definition capacitive touch display
Waveform update rate > 1 million waveforms per second
System bandwidth with N2894A standard
passive probe
200 MHz 200 MHz 350 MHz 350 MHz 500 MHz 500 MHz 700 MHz 700 MHz
System analog channels
Hardware bandwidth limits Approximately 20 MHz (selectable)
Input coupling AC, DC
Input impedance Selectable: 1 MΩ ± 1% (16 pF), 50 Ω ± 1.5%
Input sensitivity range 200 MHz ~ 500 MHz models: 1 mV/div to 5 V/div
2
(1 MΩ and 50 Ω)
1 and 1.5 GHz models: 1 mV/div to 5 V/div
2
(1 MΩ), 1 mV/div to 1 V/div (50 Ω)
Vertical resolution 8 bits (measurement resolution is 12 bits with averaging)
Maximum input voltage 1 MΩ 135 Vrms; 190 Vpk
Probing technology allows testing of higher voltages. For example, the included N2894A
10:1 probe supports testing up to 300Vrms
Use this instrument only for measurements within its specified measurement category (not
rated for CAT II, III, IV). No transient overvoltage allowed
50 Ω 50 Ω: ≤ 5 Vrms max
DC vertical gain accuracy
1
± 2.0% full scale
2
DC vertical offset accuracy ± 0.1 div ± 2 mV ± 1% of offset setting
Channel-to-channel isolation 200 MHz~1 GHz ≥ 40 dB from DC to maximum specified bandwidth of each model
1.5 GHz ≥ 40 dB from DC to 1 GHz, ≥ 35 dB from 1 to 1.5 GHz
Offset range ± 5 V (< 10 mV/div), ± 20 V (10 to 200 mV/div), ± 75 V (> 200 mV/div)
Vertical system digital channels
Digital input channels 16 digital (D0 to D15. Pod 1: D7 ~ D0, Pod 2: D15 ~ D8)
Thresholds Threshold per pod
Threshold selections TTL (+1.4 V), 5 V CMOS (+2.5 V), ECL (–1.3 V), user-defined (selectable by pod)
User-defined threshold range ± 8.0 V in 10 mV steps
Maximum input voltage ± 40 V peak CAT I
Threshold accuracy
1
± (100 mV + 3% of threshold setting)
Maximum input dynamic range ± 10 V about threshold
Minimum voltage swing 500 mVpp
Input impedance 100 kΩ ± 2% at probe tip
Input capacitance ~8 pF
Vertical resolution 1 bit
1� Denotes warranted specifications, all others are typical� Specifications are valid after a 30-minute warm-up period and ± 10 °C from firmware calibration
temperature�
2� 1 mV/div and 2 mV/div is a magnification of 4 mV/div setting� For vertical accuracy calculations, use full scale of 32 mV for 1 mV/div and 2 mV/div sensitivity
setting�
3� 1�5 GHz real time bandwidth in half-channel mode or full channel equivalent time mode�
Oscilloscope Experience Redefined:
InfiniiVision 4000 X-Series Performance Characteristics