Datasheet

Page 26
Find us at www.keysight.com
Performance Characteristics (Continued)
WaveGen – Built-in function/arbitrary waveform generator (specifications are typical) (Continued)
Frequency Sine wave and ramp accuracy:
130 ppm (frequency < 10 kHz)
50 ppm (frequency > 10 kHz)
Square wave and pulse accuracy:
[50+frequency/200] ppm (frequency < 25 kHz)
50 ppm (frequency ≥ 25 kHz)
Resolution: 0.1 Hz or 4 digits, whichever is larger
Amplitude Range:
20 mVpp to 5 Vpp into Hi-Z
2
10 mVpp to 2.5 Vpp into 50 ohms
2
Resolution: 100 μV or 3 digits, whichever is higher
Accuracy: 2% (frequency = 1 kHz)
DC offset Range:
± 2.5 V into Hi-Z
2
± 1.25 V into 50 ohms
2
Resolution: 100 μV or 3 digits, whichever is higher
Accuracy (waveform modes): ± 1.5% of offset setting ± 1% of amplitude ± 1 mV
Accuracy (DC mode): ± 1.5% of offset setting ± 3 mV
Trigger output Trigger output available on Trig out BNC
Main Output Impedance: 50 ohms typical
Isolation: Not available, main output BNC is grounded
Protection: Overload automatically disables output
Digital voltmeter (specifications are typical)
Functions ACrms, DC, DCrms, Frequency
Resolution ACV/DCV: 3 digits Frequency: 5.5 digits
Measuring rate 100 times/second
Autoranging Automatic adjustment of vertical amplification to maximize the dynamic range of measurements
Range meter Graphical display of most recent measurement, plus extrema over the previous 3 seconds
Connectivity
Standard ports One USB 2.0 hi-speed device port on rear panel. Supports USBTMC protocol
Two USB 2.0 hi-speed host ports, front and rear panel
Supports memory devices, printers and keyboards
Optional ports GPIB, LAN (10/100Base-T), WVGA video out
Trigger out BNC connector on the rear panel. Supported modes: triggers, mask, and waveform generator sync pulse
1� Gaussian Pulse: 4 Vpp maximum into Hi-Z; 2 Vpp maximum into 50 ohms
2� Sinc, Cardiac and Gaussian Pulse: ± 1�25 V into Hi-Z; ± 625 mV into 50 ohms�
3� Full resolution is not available at output due to internal attenuator stepping�