Vertical system analog channels
Hardware bandwidth limits Approximately 20 MHz (selectable)
Input coupling AC, DC
Input impedance Selectable: 1 MΩ ± 1% (14 pF), 50 Ω ± 1.5%
Input sensitivity range 100 MHz ~ 500 MHz models: 1 mV/div to 5 V/div2 (1 MΩ and 50 Ω)
1 GHz models: 1 mV/div to 5 V/div2 (1 MΩ), 1 mV/div to 1 V/div (50 Ω)
Vertical resolution 8 bits (measurement resolution is 12 bits with averaging)
Maximum input voltage 135 Vrms; 190 Vpk
Probing technology allows testing of higher voltages. For example, the included n2841A, N2842A or N2843A 10:1
probe supports testing up to 300Vrms
Use this instrument only for measurements within its specified measurement category (not rated for CAT II, III, IV).
No transient overvoltage allowed
DC vertical accuracy
DC vertical gain accuracy
1
± 2.0% full scale
2
DC vertical offset accuracy ± 0.1 div ± 2 mV ± 1% of offset setting
Channel-to-channel isolation > 100:1 from DC to maximum specified bandwidth of each model (measured with same V/div and coupling on
channels)
Offset range ± 2 V (1 mV/div to 200 mV/div)
± 50 V (> 200 mV/div to 5 V/div)
Vertical system digital channels
Digital input channels 16 digital (D0 to D15. pod 1: D7 ~ D0, Pod 2: D15 ~ D8)
Thresholds Threshold per pod
Threshold selections TTL (+1.4 V), 5V CMOS (+2.5 V), ECL (–1.3 V), user-defined (selectable by pod)
User-defined threshold range ± 8.0 V in 10 mV steps
Maximum input voltage ± 40 V peak CAT I
Threshold accuracy
1
± (100 mV + 3% of threshold setting)
Maximum input dynamic range ± 10 V about threshold
Minimum voltage swing 500 mVpp
Input impedance 100 kΩ ± 2% at probe tip
Input capacitance ~8 pF
Vertical resolution 1 bit
1� Denotes warranted specifications, all others are typical� Specifications are valid after a 30-minute warm-up period and ± 10 °C from firmware calibration
temperature�
2� 1 mV/div and 2 mV/div are a magnification of 4 mV/div setting� For vertical accuracy calculations, use full scale of 32 mV for 1 mV/div and 2 mV/div
sensitivity setting�
Performance Characteristics
DSO and MSO 3000 X-Series oscilloscopes
3000 X-Series specification overview
3012A 3014A 3024A 3032A 3034A 3052A 3054A 3102A 3104A
Bandwidth
1
(–3 dB) 100 MHz 200 MHz 350 MHz 500 MHz 1 GHz
Calculated rise time (10 to 90%) ≤ 3.5 ns ≤ 1.75 ns ≤ 1 ns ≤ 700 ps ≤ 450 ps
Input channels DSOX 2 4 4 2 4 2 4 2 4
MSOX 2 + 16 4 + 16 4 + 16 2 + 16 4 + 16 2 + 16 4 + 16 2 + 16 4 + 16
Maximum sample rate 4 GSa/s half channel, 2 GSa/s all channel 5 GSa/s half ch,
2.5 GSa/s all ch
Maximum memory depth Standard 2 Mpts, optional 4 Mpts and optional segment memory
Display size and type 8.5-inch WVGA display
Waveform update rate > 1 million waveforms per second
Number of active probes supported In general, one for 2-channel models and two for 4-channel models. Contact Keysight for specific configurations.