Datasheet
17 | Keysight | Serial Bus Options for InfiniiVision X-Series Oscilloscopes - Data Sheet
LIN
LIN input source Analog channels 1, 2, 3 or 4
Digital channels D0 to D15 except 1000 and 2000 X-Series
LIN standards LIN 1.3 or LIN 2.X
Baud rates 2400 b/s to 625 kb/s
Triggering Sync break
Frame ID (0X00HEX to 0X3FHEX)
Frame ID and data
Parity error
Checksum error
Hardware-based decode Frame ID (6-bit hex digits in yellow)
Frame ID and optional parity bits (8-bit hex digits in yellow if valid, red if parity bit error)
Data bytes (hex digits in white)
Check sum (hex digits in blue = valid, hex digits in red = error)
Sync error (“SYNC” in red)
THeader-max (“THM” in red)
TFrame-max (“TFM” in red)
Parity error (“PAR” in red)
LIN 1.3 wake-up error (“WUP” in red)
Idle bus (mid-level bus trace in dark blue)
Active bus (bi-level bus trace in dark blue)
Symbolic triggering (except 1000X and 2000X)
(based on .ldf file)
Message names
Message and signal values/encoded states
Multi-bus analysis LIN plus one other serial bus, including another LIN bus. (except 1000X and 2000X)
Specifications/Characteristics (Continued)
Tabular protocol lister display is not available on 1000 X-Series oscilloscopes.