Datasheet

15 | Keysight | Serial Bus Options for InfiniiVision X-Series Oscilloscopes - Data Sheet
Specifications/Characteristics (Continued)
CAN (1000 X-, 2000 X-, and 3000A X-Series only)
CAN input source Analog channels 1, 2, 3 or 4
Digital channels D0 to D15 non-differential. (3000 X-Series only)
Signal types Rx, Tx, CAN_L, CAN_H, Diff (L-H), Diff (H-L)
Baud rates 10 kb/s up to 5 Mb/s
Triggering Start-of-frame (SOF)
Remote frame ID (RMT)
Data frame ID (~RMT)
Remote or data frame ID
Data frame ID and data
Error frame
All errors (includes protocol “form” errors that may not generate flagged error frames)
Acknowledge errors
Overload frames
ID length: 11 bits or 29 bits (extended)
Hardware-based decode Frame ID (hex digits in yellow)
Remote frame (RMT in green)
Data length code (DLC in blue)
Data bytes (hex digits in white)
CRC (hex digits in blue = valid, hex digits in red = error)
Error frame (bi-level bus trace and ERR message in red)
Form error (bi-level bus trace and “?” in red)
Overload frame (“OVRLD” in blue)
Idle bus (mid-level bus trace in dark blue)
Active bus (bi-level bus trace in dark blue)
Multi-bus analysis CAN plus one other serial bus, including another CAN bus. (3000 X-Series only)
Totalize function Total frames, total overload frames, total error frames, bus utilization (bus load)
Eye-diagram mask testing (3000 X-Series
requires DSOX3MASK, 1000 X-Series requires
DSO model)
Various downloadable mask files available based on differential probing polarity, baud rate and network
length
Tabular protocol lister display is not available on 1000 X-Series oscilloscopes.