User Manual

Table Of Contents
94 Keysight InfiniiVision 1000 X-Series Oscilloscopes Programmer's Guide
4 Commands Quick Reference
:TRIGger:NREJect {{0
| OFF} | {1 | ON}}
(see page 581)
:TRIGger:NREJect?
(see page 581)
{0 | 1}
:TRIGger:SWEep
<sweep> (see page 582)
:TRIGger:SWEep? (see
page 582)
<sweep> ::= {AUTO | NORMal}
Table 29 General :TRIGger Commands Summary (continued)
Command Query Options and Query Returns
Table 30 :TRIGger[:EDGE] Commands Summary
Command Query Options and Query Returns
:TRIGger[:EDGE]:COUPl
ing {AC | DC |
LFReject} (see
page 584)
:TRIGger[:EDGE]:COUPl
ing? (see page 584)
{AC | DC | LFReject}
:TRIGger[:EDGE]:LEVel
<level> [,<source>]
(see page 585)
:TRIGger[:EDGE]:LEVel
? [<source>] (see
page 585)
For internal triggers, <level>
::= .75 x full-scale voltage from
center screen in NR3 format.
For external triggers, <level>
::= ±(external range setting) in
NR3 format.
<source> ::= {CHANnel<n> |
EXTernal}
<n> ::= 1 to (# analog channels)
in NR1 format
:TRIGger[:EDGE]:REJec
t {OFF | LFReject |
HFReject} (see
page 586)
:TRIGger[:EDGE]:REJec
t? (see page 586)
{OFF | LFReject | HFReject}
:TRIGger[:EDGE]:SLOPe
<polarity> (see
page 587)
:TRIGger[:EDGE]:SLOPe
? (see page 587)
<polarity> ::= {POSitive |
NEGative | EITHer | ALTernate}
:TRIGger[:EDGE]:SOURc
e <source> (see
page 588)
:TRIGger[:EDGE]:SOURc
e? (see page 588)
<source> ::= {CHANnel<n> |
EXTernal | LINE | WGEN}
<n> ::= 1 to (# analog channels)
in NR1 format