728 Keysight InfiniiVision 1000 X-Series Oscilloscopes Programmer's Guide
31 Obsolete and Discontinued Commands
:MEASure:VSTOp
(see page 782)
Command Syntax
:MEASure:VSTOp <vstop_argument>
<vstop_argument> ::= value for Y2 cursor
The :MEASure:VSTOp command moves the vertical marker 2 (Y2 cursor) to the
specified value corresponding to the selected source. The source can be selected
by the MARKer:X2Y2source command.
Query Syntax
:MEASure:VSTOp?
The :MEASure:VSTOp? query returns the current value of the Y2 cursor.
Return Format
<value><NL>
<value> ::= value of the Y2 cursor in NR3 format
See Also "Introduction to :MARKer Commands" on page 322
"Introduction to :MEASure Commands" on page 345
":MARKer:Y1Position" on page 332
":MARKer:Y2Position" on page 333
":MARKer:YDELta" on page 334
":MARKer:X2Y2source" on page 328
":MEASure:SOURce" on page 371
":MEASure:TDELta" on page 718
":MEASure:TSTArt" on page 722
NOTE
The short form of this command, VSTO, does not follow the defined Long Form to Short Form
Truncation Rules (see page 784). The normal short form, VST, would be the same for both
VSTArt and VSTOp, so sending VST for the VSTOp command produces an error.
NOTE
The :MEASure:VSTOp command is an obsolete command provided for compatibility to
previous oscilloscopes. Use the :MARKer:Y2Position command (see page 333) instead.