:MEASure:VSTArt
(see page 782)
Command Syntax
:MEASure:VSTArt <vstart_argument>
<vstart_argument> ::= value for vertical marker 1
The :MEASure:VSTArt command moves the vertical marker (Y1 cursor) to the
specified value corresponding to the selected source. The source can be selected
by the MARKer:X1Y1source command.
Query Syntax
:MEASure:VSTArt?
The :MEASure:VSTArt? query returns the current value of the Y1 cursor.
Return Format
<value><NL>
<value> ::= voltage at voltage marker 1 in NR3 format
See Also "Introduction to :MARKer Commands" on page 322
"Introduction to :MEASure Commands" on page 345
":MARKer:Y1Position" on page 332
":MARKer:Y2Position" on page 333
":MARKer:YDELta" on page 334
":MARKer:X1Y1source" on page 326
":MEASure:SOURce" on page 371
":MEASure:TDELta" on page 718
":MEASure:TSTArt" on page 722
NOTE
The short form of this command, VSTA, does not follow the defined Long Form to Short Form
Truncation Rules (see page 784). The normal short form, VST, would be the same for both
VSTArt and VSTOp, so sending VST for the VSTArt command produces an error.
NOTE
The :MEASure:VSTArt command is an obsolete command provided for compatibility to
previous oscilloscopes. Use the :MARKer:Y1Position command (see page 332) instead.