:MEASure:TSTOp
(see page 782)
Command Syntax
:MEASure:TSTOp <value> [suffix]
<value> ::= time at the stop marker in seconds
[suffix] ::= {s | ms | us | ns | ps}
The :MEASure:TSTOp command moves the stop marker (X2 cursor) to the
specified time with respect to the trigger time.
Query Syntax
:MEASure:TSTOp?
The :MEASure:TSTOp? query returns the time at the stop marker (X2 cursor).
Return Format
<value><NL>
<value> ::= time at the stop marker in NR3 format
See Also "Introduction to :MARKer Commands" on page 322
"Introduction to :MEASure Commands" on page 345
":MARKer:X1Position" on page 325
":MARKer:X2Position" on page 327
":MARKer:XDELta" on page 329
":MEASure:TDELta" on page 718
":MEASure:TSTArt" on page 722
NOTE
The short form of this command, TSTO, does not follow the defined Long Form to Short Form
Truncation Rules (see page 784). The normal short form "TST" would be the same for both
TSTArt and TSTOp, so sending TST for the TSTOp command produces an error.
NOTE
The :MEASure:TSTOp command is an obsolete command provided for compatibility to
previous oscilloscopes. Use the :MARKer:X2Position command (see page 327) instead.