722 Keysight InfiniiVision 1000 X-Series Oscilloscopes Programmer's Guide
31 Obsolete and Discontinued Commands
:MEASure:TSTArt
(see page 782)
Command Syntax
:MEASure:TSTArt <value> [suffix]
<value> ::= time at the start marker in seconds
[suffix] ::= {s | ms | us | ns | ps}
The :MEASure:TSTArt command moves the start marker (X1 cursor) to the
specified time with respect to the trigger time.
Query Syntax
:MEASure:TSTArt?
The :MEASure:TSTArt? query returns the time at the start marker (X1 cursor).
Return Format
<value><NL>
<value> ::= time at the start marker in NR3 format
See Also "Introduction to :MARKer Commands" on page 322
"Introduction to :MEASure Commands" on page 345
":MARKer:X1Position" on page 325
":MARKer:X2Position" on page 327
":MARKer:XDELta" on page 329
":MEASure:TDELta" on page 718
":MEASure:TSTOp" on page 723
NOTE
The short form of this command, TSTA, does not follow the defined Long Form to Short Form
Truncation Rules (see page 784). The normal short form "TST" would be the same for both
TSTArt and TSTOp, so sending TST for the TSTArt command produces an error.
NOTE
The :MEASure:TSTArt command is an obsolete command provided for compatibility to
previous oscilloscopes. Use the :MARKer:X1Position command (see page 325) instead.