:MEASure:TMIN
(see page 782)
Command Syntax
:MEASure:TMIN [<source>]
<source> ::= {CHANnel<n> | FUNCtion | MATH}
<n> ::= 1 to (# analog channels) in NR1 format
The :MEASure:TMIN command installs a screen measurement and starts an
X-at-Min-Y measurement on the selected waveform. If the optional source is
specified, the current source is modified.
Query Syntax
:MEASure:TMIN? [<source>]
The :MEASure:TMIN? query returns the horizontal axis value at which the minimum
vertical value occurs on the current source. If the optional source is specified, the
current source is modified. If all channels are off, the query returns 9.9E+37.
Return Format
<value><NL>
<value> ::= time at minimum in NR3 format
See Also "Introduction to :MEASure Commands" on page 345
":MEASure:TMAX" on page 720
":MEASure:XMAX" on page 387
":MEASure:XMIN" on page 388
NOTE
The :MEASure:TMIN command is obsolete and is provided for backward compatibility to
previous oscilloscopes. Use the :MEASure:XMIN command (see page 388) instead.