:WGEN Commands 29
Keysight InfiniiVision 1000 X-Series Oscilloscopes Programmer's Guide 665
:WGEN:MODulation:AM:DEPTh
(see page 782)
Command Syntax
:WGEN:MODulation:AM:DEPTh <percent>
<percent> ::= AM depth percentage from 0% to 100% in NR1 format
The :WGEN:MODulation:AM:DEPTh command specifies the amount of amplitude
modulation.
AM Depth refers to the portion of the amplitude range that will be used by the
modulation. For example, a depth setting of 80% causes the output amplitude to
vary from 10% to 90% (90% – 10% = 80%) of the original amplitude as the
modulating signal goes from its minimum to maximum amplitude.
Query Syntax
:WGEN:MODulation:AM:DEPTh?
The :WGEN:MODulation:AM:DEPTh? query returns the AM depth percentage
setting.
Return Format
<percent><NL>
<percent> ::= AM depth percentage from 0% to 100% in NR1 format
See Also ":WGEN:MODulation:AM:FREQuency" on page 666
":WGEN:MODulation:FM:DEViation" on page 667
":WGEN:MODulation:FM:FREQuency" on page 668
":WGEN:MODulation:FSKey:FREQuency" on page 669
":WGEN:MODulation:FSKey:RATE" on page 670
":WGEN:MODulation:FUNCtion" on page 671
":WGEN:MODulation:FUNCtion:RAMP:SYMMetry" on page 672
":WGEN:MODulation:STATe" on page 674
":WGEN:MODulation:TYPE" on page 675