660 Keysight InfiniiVision 1000 X-Series Oscilloscopes Programmer's Guide
29 :WGEN Commands
For all waveform types, the output amplitude, into 50 Ω, can be adjusted from
10 mVpp to 2.5 Vpp (or from 20 mVpp to 5 Vpp into and open-circuit load).
Query Syntax
:WGEN:FUNCtion?
The :WGEN:FUNCtion? query returns the currently selected signal type.
Return Format
<signal><NL>
<signal> ::= {SIN | SQU | RAMP | PULS | NOIS | DC}
RAMP Use these commands to set the ramp signal parameters:
":WGEN:FREQuency" on page 658
":WGEN:PERiod" on page 680
":WGEN:VOLTage" on page 682
":WGEN:VOLTage:OFFSet" on page 685
":WGEN:VOLTage:HIGH" on page 683
":WGEN:VOLTage:LOW" on page 684
":WGEN:FUNCtion:RAMP:SYMMetry" on page 663
The frequency can be adjusted from 100 mHz to 100 kHz.
Symmetry represents the amount of time per cycle that the ramp waveform is rising
and can be adjusted from 0% to 100%.
PULSe Use these commands to set the pulse signal parameters:
":WGEN:FREQuency" on page 658
":WGEN:PERiod" on page 680
":WGEN:VOLTage" on page 682
":WGEN:VOLTage:OFFSet" on page 685
":WGEN:VOLTage:HIGH" on page 683
":WGEN:VOLTage:LOW" on page 684
":WGEN:FUNCtion:PULSe:WIDTh" on page 662
The frequency can be adjusted from 100 mHz to 10 MHz.
The pulse width can be adjusted from 20 ns to the period minus 20 ns.
DC Use this command to set the DC level:
":WGEN:VOLTage:OFFSet" on page 685
NOISe Use these commands to set the noise signal parameters:
":WGEN:VOLTage" on page 682
":WGEN:VOLTage:OFFSet" on page 685
":WGEN:VOLTage:HIGH" on page 683
":WGEN:VOLTage:LOW" on page 684
Waveform
Type
Characteristics