User Manual

Table Of Contents
:WGEN Commands 29
Keysight InfiniiVision 1000 X-Series Oscilloscopes Programmer's Guide 659
:WGEN:FUNCtion
(see page 782)
Command Syntax
:WGEN:FUNCtion <signal>
<signal> ::= {SINusoid | SQUare | RAMP | PULSe | NOISe | DC}
The :WGEN:FUNCtion command selects the type of waveform:
Waveform
Type
Characteristics
SINusoid Use these commands to set the sine signal parameters:
":WGEN:FREQuency" on page 658
":WGEN:PERiod" on page 680
":WGEN:VOLTage" on page 682
":WGEN:VOLTage:OFFSet" on page 685
":WGEN:VOLTage:HIGH" on page 683
":WGEN:VOLTage:LOW" on page 684
The frequency can be adjusted from 100 mHz to 20 MHz.
SQUare Use these commands to set the square wave signal parameters:
":WGEN:FREQuency" on page 658
":WGEN:PERiod" on page 680
":WGEN:VOLTage" on page 682
":WGEN:VOLTage:OFFSet" on page 685
":WGEN:VOLTage:HIGH" on page 683
":WGEN:VOLTage:LOW" on page 684
":WGEN:FUNCtion:SQUare:DCYCle" on page 664
The frequency can be adjusted from 100 mHz to 10 MHz.
The duty cycle can be adjusted from 20% to 80%.