User Manual

Table Of Contents
568 Keysight InfiniiVision 1000 X-Series Oscilloscopes Programmer's Guide
26 :TIMebase Commands
:TIMebase:WINDow:RANGe
(see page 782)
Command Syntax
:TIMebase:WINDow:RANGe <range value>
<range value> ::= range value in seconds in NR3 format
The :TIMebase:WINDow:RANGe command sets the full-scale horizontal time in
seconds for the zoomed (delayed) window. The range is 10 times the current
zoomed view window seconds per division setting. The main sweep range
determines the range for this command. The maximum value is one half of the
:TIMebase:RANGe value.
Query Syntax
:TIMebase:WINDow:RANGe?
The :TIMebase:WINDow:RANGe? query returns the current window timebase
range setting.
Return Format
<value><NL>
<value> ::= range value in seconds
See Also "Introduction to :TIMebase Commands" on page 560
":TIMebase:RANGe" on page 563
":TIMebase:POSition" on page 562
":TIMebase:SCALe" on page 565