:SBUS<n>:SPI:TRIGger:PATTern:MISO:DATA
(see page 782)
Command Syntax
:SBUS<n>:SPI:TRIGger:PATTern:MISO:DATA <string>
<string> ::= "nn...n" where n ::= {0 |1|X|$}
<string ::= "0xnn...n" where n ::= {0,..,9 | A,..,F | X | $}
The :SBUS<n>:SPI:TRIGger:PATTern:MISO:DATA command defines the SPI data
pattern resource according to the string parameter. This pattern, along with the
data width, control the data pattern searched for in the data stream.
If the string parameter starts with "0x", it is a hexadecimal string made up of
hexadecimal and X (don't care) characters; otherwise, it is a binary string made up
of 0, 1, and X (don't care) characters.
Query Syntax
:SBUS<n>:SPI:TRIGger:PATTern:MISO:DATA?
The :SBUS<n>:SPI:TRIGger:PATTern:MISO:DATA? query returns the current
settings of the specified SPI data pattern resource in the binary string format.
Return Format
<string><NL>
See Also "Introduction to :TRIGger Commands" on page 571
":SBUS<n>:SPI:TRIGger:PATTern:MISO:WIDTh" on page 508
":SBUS<n>:SPI:SOURce:MISO" on page 505
NOTE
The :SBUS<n>:SPI:TRIGger:PATTern:MISO:WIDTh should be set before
:SBUS<n>:SPI:TRIGger:PATTern:MISO:DATA.