User Manual

Table Of Contents
:SBUS<n> Commands 24
Keysight InfiniiVision 1000 X-Series Oscilloscopes Programmer's Guide 491
:SBUS<n>:LIN:TRIGger
(see page 782)
Command Syntax
:SBUS<n>:LIN:TRIGger <condition>
<condition> ::= {SYNCbreak | ID | DATA}
The :SBUS<n>:LIN:TRIGger command sets the LIN trigger condition to be:
•SYNCbreak Sync Break.
•ID Frame ID.
Use the :SBUS<n>:LIN:TRIGger:ID command to specify the frame ID.
•DATA Frame ID and Data.
Use the :SBUS<n>:LIN:TRIGger:ID command to specify the frame ID.
Use the :SBUS<n>:LIN:TRIGger:PATTern:DATA:LENGth and
:SBUS<n>:LIN:TRIGger:PATTern:DATA commands to specify the data string
length and value.
Query Syntax
:SBUS<n>:LIN:TRIGger?
The :SBUS<n>:LIN:TRIGger? query returns the current LIN trigger value.
Return Format
<condition><NL>
<condition> ::= {SYNC | ID | DATA}
Errors "-241, Hardware missing" on page 745
See Also "Introduction to :TRIGger Commands" on page 571
":TRIGger:MODE" on page 580
":SBUS<n>:LIN:TRIGger:ID" on page 492
":SBUS<n>:LIN:TRIGger:PATTern:DATA:LENGth" on page 495
":SBUS<n>:LIN:TRIGger:PATTern:DATA" on page 493
":SBUS<n>:LIN:SOURce" on page 488