User Manual

Table Of Contents
486 Keysight InfiniiVision 1000 X-Series Oscilloscopes Programmer's Guide
24 :SBUS<n> Commands
:SBUS<n>:LIN:SAMPlepoint
(see page 782)
Command Syntax
:SBUS<n>:LIN:SAMPlepoint <value>
<value><NL>
<value> ::= {60 | 62.5 | 68 | 70 | 75 | 80 | 87.5} in NR3 format
The :SBUS<n>:LIN:SAMPlepoint command sets the point during the bit time
where the bit level is sampled to determine whether the bit is dominant or
recessive. The sample point represents the percentage of time between the
beginning of the bit time to the end of the bit time.
Query Syntax
:SBUS<n>:LIN:SAMPlepoint?
The :SBUS<n>:LIN:SAMPlepoint? query returns the current LIN sample point
setting.
Return Format
<value><NL>
<value> ::= {60 | 62.5 | 68 | 70 | 75 | 80 | 87.5} in NR3 format
See Also "Introduction to :TRIGger Commands" on page 571
":TRIGger:MODE" on page 580
":SBUS<n>:LIN:TRIGger" on page 491
NOTE
The sample point values are not limited by the baud rate.