User Manual

Table Of Contents
:MTESt Commands 21
Keysight InfiniiVision 1000 X-Series Oscilloscopes Programmer's Guide 411
:MTESt:RMODe:FACTion:SAVE
(see page 782)
Command Syntax
:MTESt:RMODe:FACTion:SAVE <on_off>
<on_off> ::= {{1 | ON} | {0 | OFF}}
The :MTESt:RMODe:FACTion:SAVE command sets saving on mask failures on or
off.
See Chapter 23, “:SAVE Commands,” starting on page 431 for more information
on save options.
Query Syntax
:MTESt:RMODe:FACTion:SAVE?
The :MTESt:RMODe:FACTion:SAVE? query returns the current mask failure save
setting.
Return Format
<on_off><NL>
<on_off> ::= {1 | 0}
See Also "Introduction to :MTESt Commands" on page 391
":MTESt:RMODe:FACTion:MEASure" on page 409
":MTESt:RMODe:FACTion:PRINt" on page 410
":MTESt:RMODe:FACTion:STOP" on page 412
NOTE
Setting :MTESt:RMODe:FACTion:SAVE ON automatically sets :MTESt:RMODe:FACTion:PRINt
OFF.