Keysight InfiniiVision 1000 X-Series Oscilloscopes Programmer's Guide 377
:MEASure:VAMPlitude
(see page 782)
Command Syntax
:MEASure:VAMPlitude [<source>]
<source> ::= {CHANnel<n> | FUNCtion | MATH | WMEMory<r>}
<n> ::= 1 to (# analog channels) in NR1 format
<r> ::= 1-2 in NR1 format
The :MEASure:VAMPlitude command installs a screen measurement and starts a
vertical amplitude measurement. If the optional source parameter is specified, the
current source is modified.
Query Syntax
:MEASure:VAMPlitude? [<source>]
The :MEASure:VAMPlitude? query measures and returns the vertical amplitude of
the waveform. To determine the amplitude, the instrument measures Vtop and
Vbase, then calculates the amplitude as follows:
vertical amplitude = Vtop - Vbase
Return Format
<value><NL>
<value> ::= the amplitude of the selected waveform in NR3 format
See Also "Introduction to :MEASure Commands" on page 345
":MEASure:SOURce" on page 371
":MEASure:VBASe" on page 379
":MEASure:VTOP" on page 385
":MEASure:VPP" on page 382