User Manual

Table Of Contents
:MEASure Commands 20
Keysight InfiniiVision 1000 X-Series Oscilloscopes Programmer's Guide 363
:MEASure:PERiod
(see page 782)
Command Syntax
:MEASure:PERiod [<source>]
<source> ::= {CHANnel<n> | FUNCtion | MATH | WMEMory<r> | EXTernal}
<n> ::= 1 to (# of analog channels) in NR1 format
<r> ::= 1-2 in NR1 format
The :MEASure:PERiod command installs a screen measurement and starts the
period measurement. If the optional source parameter is specified, the current
source is modified.
Query Syntax
:MEASure:PERiod? [<source>]
The :MEASure:PERiod? query measures and outputs the period of the cycle closest
to the trigger reference on the screen. The period is measured at the midpoint of
the upper and lower thresholds.
IF the edge closest to the trigger reference on screen is rising:
THEN period = (time at trailing rising edge - time at leading rising edge)
ELSE period = (time at trailing falling edge - time at leading falling edge)
Return Format
<value><NL>
<value> ::= waveform period in seconds in NR3 format
See Also "Introduction to :MEASure Commands" on page 345
":MEASure:SOURce" on page 371
":MEASure:NWIDth" on page 360
":MEASure:PWIDth" on page 366
":MEASure:FREQuency" on page 358
Example Code "Example Code" on page 371
NOTE
This command is not available if the source is FFT (Fast Fourier Transform).