360 Keysight InfiniiVision 1000 X-Series Oscilloscopes Programmer's Guide
20 :MEASure Commands
:MEASure:NWIDth
(see page 782)
Command Syntax
:MEASure:NWIDth [<source>]
<source> ::= {CHANnel<n> | FUNCtion | MATH | WMEMory<r> | EXTernal}
<n> ::= 1 to (# of analog channels) in NR1 format
<r> ::= 1-2 in NR1 format
The :MEASure:NWIDth command installs a screen measurement and starts a
negative pulse width measurement. If the optional source parameter is not
specified, the current source is modified.
Query Syntax
:MEASure:NWIDth? [<source>]
The :MEASure:NWIDth? query measures and outputs the width of the negative
pulse on the screen closest to the trigger reference using the midpoint between
the upper and lower thresholds.
FOR the negative pulse closest to the trigger point:
width = (time at trailing rising edge - time at leading falling edge)
Return Format
<value><NL>
<value> ::= negative pulse width in seconds in NR3 format
See Also "Introduction to :MEASure Commands" on page 345
":MEASure:SOURce" on page 371
":MEASure:PWIDth" on page 366
":MEASure:PERiod" on page 363
NOTE
This command is not available if the source is FFT (Fast Fourier Transform).